DOI 10.1097/mao.0000000000004897

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026

Effect of Pulse Phase Duration on the Electrically Evoked Compound Action Potentials in Pediatric Cochlear Implant Users.

Saravanan Palani, Antony P Jawahar, Megha, Geetha Chinnaraj, Manjula Puttabasappa

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

The present study examined the effect of pulse phase duration on electrically evoked compound action potentials (eCAP) measurability, thresholds, and their relationship with behavioral comfort levels on implant models with default pulse phase duration of 37 µs.

STUDY

DESIGN

Within-subject comparison of prospectively collected clinical data.

SETTING

Large cochlear implant (CI) program at government funded academic and research institute.

PATIENTS: Prelingually deafened paediatric unilateral CI users using Cochlear Nucleus CI422 implant with slim-straight electrode array (n = 62).

MAIN OUTCOME MEASURES: eCAPs were recorded across 5 electrodes across the array (E01, E06, E11, E16, and E22) in 62 paediatric CI422 users using 25 µs (via AutoNRT) and 37 µs (via advanced-NRT) durations.

RESULTS

The χ 2 test revealed significantly higher eCAP measurability with 37 µs than with 25 µs pulse phase duration. The 37 µs eCAP thresholds fell within 62% to 73% of the electrical dynamic range, whereas the 25 µs thresholds exceeded it (105% to 117%). A moderate positive correlation (r = 0.40 to 0.6) was found between eCAP thresholds and behavioral comfort levels for both pulse phase durations. The comparison of eCAP thresholds in terms of clinical units (CL) revealed significantly lower eCAP thresholds with 37 µs pulse phase duration ( P < 0.05), whereas, in terms of charge levels (nC), there were no significant differences in thresholds measured with both 25 and 37 µs pulse phase duration in all the tested electrodes ( P > 0.05).

CONCLUSION

eCAP measurements with 37 µs pulse phase duration can maximize eCAP recordings, especially for Cochlear Nucleus implants with slim-straight electrode arrays.

Для цитирования: Saravanan Palani, Antony P Jawahar, Megha, Geetha Chinnaraj, Manjula Puttabasappa. Effect of Pulse Phase Duration on the Electrically Evoked Compound Action Potentials in Pediatric Cochlear Implant Users.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2026. DOI: 10.1097/mao.0000000000004897.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.