Corneal Vibrations during Intraocular Pressure Measurement with an Air-Puff Method.

Koprowski Robert, Wilczyński Sławomir

СтатьяОбзорПеревод в очереди

Аннотация

INTRODUCTION

The paper presents a commentary on the method of analysis of corneal vibrations occurring during eye pressure measurements with air-puff tonometers, for example, Corvis. The presented definition and measurement method allow for the analysis of image sequences of eye responses-cornea deformation. In particular, the outer corneal contour and sclera fragments are analysed, and 3D reconstruction is performed.

METHODS

On this basis, well-known parameters such as eyeball reaction or corneal response are determined. The next steps of analysis allow for automatic and reproducible separation of four different corneal vibrations. These vibrations are associated with (1) the location of the maximum of cornea deformation; (2) the cutoff area measured in relation to the cornea in a steady state; (3) the maximum of peaks occurring between applanations; and (4) the other characteristic points of the corneal contour.

RESULTS

The results obtained enable (1) automatic determination of the amplitude of vibrations; (2) determination of the frequency of vibrations; and (3) determination of the correlation between the selected types of vibrations.

CONCLUSIONS

These are diagnostic features that can be directly applied clinically for new and archived data.

Для цитирования: Koprowski Robert, Wilczyński Sławomir. Corneal Vibrations during Intraocular Pressure Measurement with an Air-Puff Method.. Journal of healthcare engineering. 2018. DOI: 10.1155/2018/5705749.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.