DOI 10.1016/j.athoracsur.2017.07.013

Медицинская библиотека · The Annals of thoracic surgery · 2017

Medium-Term Outcomes After Implantation of Expanded Polytetrafluoroethylene Valved Conduit.

Ootaki Yoshio, Welch Allison S, Walsh Michael J, Quartermain Michael D, Williams Derek A, Ungerleider Ross M

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Pulmonary valve replacement is becoming the most frequent congenital heart surgery performed on adolescents and young adults. Numerous surgical options are available, including autologous pericardium, mechanical valves, allografts, and bioprosthetic valves. Each option has limitations with durability and freedom from reintervention for stenosis or insufficiency, particularly in the pediatric population. The purpose of this study was to analyze our uniquely designed, expanded polytetrafluoroethylene (ePTFE) valved conduit (VC) for right ventricular outflow tract reconstruction.

METHODS

Beginning in 2012, ePTFE VC were implanted in 26 patients with a median age of 9.8 years (range, 1.0 to 15.9). Bicuspid VC were used in the first 3 patients, and tricuspid VC were used in 23 patients. The ePTFE VC is fashioned from commercially available ePTFE tube graft (16 mm in 1 patient, 20 mm in 7 patients, 24 mm in 18 patients) and 0.1 mm thick ePTFE membrane for the leaflet material. Valve function was assessed by echocardiogram after the implantation.

RESULTS

There were no postoperative hospital deaths. Mean follow-up was 2.6 years (range, 4 months to 4.7 years), and there have been no reoperations for the ePTFE VC. There was one reintervention for stenosis at the distal anastomosis. Pulmonary insufficiency was mild or less in 24 patients (92%). The average peak instantaneous pressure gradient between the right ventricle and the pulmonary artery was 22.4 ± 15.1 mm Hg at latest follow-up.

CONCLUSIONS

Compared with historical data for other pulmonary valve replacement options, our ePTFE VC shows superior medium-term performance, with less reintervention or significant valve dysfunction.

Для цитирования: Ootaki Yoshio, Welch Allison S, Walsh Michael J, Quartermain Michael D, Williams Derek A, Ungerleider Ross M. Medium-Term Outcomes After Implantation of Expanded Polytetrafluoroethylene Valved Conduit.. The Annals of thoracic surgery. 2017. DOI: 10.1016/j.athoracsur.2017.07.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.