Circulation. Cardiovascular interventions · 2024
Transcatheter Pulmonary Valve Replacement With the Harmony Valve in Patients Who Do Not Meet Recommended Oversizing Criteria on the Screening Perimeter Plot.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
McElhinney Doff B, Gillespie Matthew J, Aboulhosn Jamil A, Cabalka Allison K, Morray Brian H, Balzer David T, Qureshi Athar M, Hoskoppal Arvind K, Goldstein Bryan H
Аннотация
BACKGROUND: Anatomic selection for Harmony valve implant is determined with the aid of a screening report and perimeter plot (PP) that depicts the perimeter-derived radius along the right ventricular outflow tract (RVOT) and projects device oversizing. The PP provides an estimation of suitability for implant, but its sensitivity as a screening method is unknown. This study was performed to describe anatomic features and outcomes in patients who underwent Harmony TPV25 implant despite a PP that predicted inadequate oversizing.
METHODS: We reviewed RVOT anatomic features and measurements in patients who underwent transcatheter pulmonary valve replacement with the Harmony TPV25 device despite a PP that predicted inadequate oversizing.
RESULTS: This study included 22 patients. There were no unsuccessful implants or adverse valve-related events. Anatomic features varied, but all patients fit into 1 of 3 anatomic types characterized by differences in RVOT dimensions. Type 1 patients (n=9) had a long RVOT with a choke point and a wide main pulmonary artery. Type 2 patients (n=6) had a short RVOT that was pyramidal in shape, with no choke point, and extensive main pulmonary artery lengthening/expansion during systole. Type 3 patients (n=7) had a short, bulbous main pulmonary artery with a choke point and an open pulmonary artery bifurcation.
CONCLUSIONS: Transcatheter pulmonary valve replacement with the Harmony valve is feasible in some patients whose PP fit analysis predicts inadequate oversizing. All cases in this series fit into 1 of 3 anatomic patterns, which are not identified in the screening report. Implanters must review cases individually to assess the feasibility of the implant.
METHODS: We reviewed RVOT anatomic features and measurements in patients who underwent transcatheter pulmonary valve replacement with the Harmony TPV25 device despite a PP that predicted inadequate oversizing.
RESULTS: This study included 22 patients. There were no unsuccessful implants or adverse valve-related events. Anatomic features varied, but all patients fit into 1 of 3 anatomic types characterized by differences in RVOT dimensions. Type 1 patients (n=9) had a long RVOT with a choke point and a wide main pulmonary artery. Type 2 patients (n=6) had a short RVOT that was pyramidal in shape, with no choke point, and extensive main pulmonary artery lengthening/expansion during systole. Type 3 patients (n=7) had a short, bulbous main pulmonary artery with a choke point and an open pulmonary artery bifurcation.
CONCLUSIONS: Transcatheter pulmonary valve replacement with the Harmony valve is feasible in some patients whose PP fit analysis predicts inadequate oversizing. All cases in this series fit into 1 of 3 anatomic patterns, which are not identified in the screening report. Implanters must review cases individually to assess the feasibility of the implant.
Предметные рубрики MeSH
HumansPulmonary ValveHeart Valve ProsthesisHeart Valve Prosthesis ImplantationCardiac CatheterizationProsthesis DesignTreatment OutcomeMaleFemaleAdolescentYoung AdultChildAdultRetrospective StudiesPredictive Value of TestsPulmonary Valve InsufficiencyPatient SelectionClinical Decision-MakingRecovery of FunctionHemodynamics