DOI 10.1007/s11547-009-0462-y

Медицинская библиотека · La Radiologia medica · 2009

Radiation dose saving through the use of cone-beam CT in hearing-impaired patients.

Faccioli N, Barillari M, Guariglia S, Zivelonghi E, Rizzotti A, Cerini R, Mucelli R Pozzi

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: Bionic ear implants provide a solution for deafness. Patients treated with these hearing devices are often children who require close follow-up with frequent functional and radiological examinations; in particular, multislice computed tomography (MSCT). Dental volumetric cone-beam CT (CBCT) has been reported as a reliable technique for acquiring images of the temporal bone while delivering low radiation doses and containing costs. The aim of this study was to assess, in terms of radiation dose and image quality, the possibility of using CBCT as an alternative to MSCT in patients with bionic ear implants.

MATERIALS AND

METHODS

One hundred patients (mean age 26 years, range 7-43) with Vibrant SoundBridge implants on the round window underwent follow-up: 85 with CBCT and 15 with MSCT. We measured the average tissue-absorbed doses during both MSCT and CBCT scans. Each scan was focused on the temporal bone with the smallest field of view and a low-dose protocol. In order to estimate image quality, we obtained data about slice thickness, high- and low-contrast resolution, uniformity and noise by using an AAPM CT performance phantom.

RESULTS

Although the CBCT images were qualitatively inferior to those of MSCT, they were sufficiently diagnostic to allow evaluation of the position of the implants. The effective dose of MSCT was almost three times higher than that of CBCT.

CONCLUSIONS

Owing to low radiation dose and sufficient image quality, CBCT could be considered an adequate technique for postoperative imaging and follow-up of patients with bionic ear implants.

Для цитирования: Faccioli N, Barillari M, Guariglia S, Zivelonghi E, Rizzotti A, Cerini R, Mucelli R Pozzi. Radiation dose saving through the use of cone-beam CT in hearing-impaired patients.. La Radiologia medica. 2009. DOI: 10.1007/s11547-009-0462-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.