AIMS: Cardiac pacing is a difficult technique in children, particularly in patients with congenital heart defects (CHDs). Few studies to date have addressed this topic.

METHODS AND

RESULTS

We performed a retrospective analysis of the results of a single centre. Between 1982 and 2008, 287 patients with CHD, median age of 5 years (25-75%, 1-11) underwent cardiac pacing for sinus node dysfunction (SND) and atrioventricular block (AVB); 97% of patients underwent at least one heart surgery. Endocardial systems (Endo) were implanted in 117 patients, epicardial systems (Epi) in 170, with 595 leads (228 Endo, 367 Epi). Endocardial systems showed a significantly older age group with more frequent SND; Epi a younger age group, with more frequent AVB, greater number of surgical interventions. Perioperative complications were mortality 0.6% (Epi), pericardial effusion 0.6% (Epi), and haemothorax 3.4% (Endo). The median follow-up is 5 (2-10) years: the pacing system failed in 29% of patients, 13% Endo, and 40% Epi (P < 0.0001). Multivariate analysis showed a significantly higher risk of failure for Epi, a lower implant age, greater the number of leads implanted. The risk of malfunction of the leads increases significantly for Epi and the younger age when implanted. The steroid-eluting leads have a lower risk of malfunction (P = 0.05), steroid-eluting Endo leads provide significantly better outcomes than Epi.

CONCLUSION

Cardiac pacing in paediatric patients with CHD shows satisfactory results in the long term. Endocardial systems show significantly better results than Epi systems. A younger age when implanted is a risk factor for complications at follow-up.

Для цитирования: Silvetti Massimo Stefano, Drago Fabrizio, Di Carlo Duccio, Placidi Silvia, Brancaccio Gianluca, Carotti Adriano. Cardiac pacing in paediatric patients with congenital heart defects: transvenous or epicardial?. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2013. DOI: 10.1093/europace/eut029.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.