Characteristics of Medial and Lateral Wall Cochlear Implant Arrays Demonstrated with Cone Beam CT and Evoked Electrical Auditory Brainstem Responses.

Alnafjan Fadwa, da Cruz Melville J, McMahon Catherine M

СтатьяПеревод в очереди

Аннотация

HYPOTHESIS: This study investigates how electrode array types-straight versus perimodiolar-affect cochlear implant (CI) placement (i.e., modiolar proximity and angular depth) and outcomes using cone beam computed tomography (CBCT) and evoked electrical auditory brainstem responses (eABR).

BACKGROUND

Cochlear implants (CIs) have revolutionized auditory rehabilitation, yet optimizing electrode placement remains crucial for improving speech perception and pitch discrimination. Array designs influence electrode proximity to neural structures, impacting neural stimulation efficacy.

METHODS

We analyzed 108 patients using CBCT to compare straight and perimodiolar electrode arrays (each with 22 electrodes) in terms of angular depth, spiral diameter, intracochlear positioning index (ICPI), electrode to modiolus, medial-lateral distance (EMML), and wrapping factor. A subset of 50 patients underwent eABR assessments across cochlear regions.

RESULTS

Significant differences were found between array types. Straight arrays showed less variability with consistent lateral placement compared to perimodiolar arrays. Measurements revealed differences: wrapping factor (M = 0.692 vs. 0.826; t(11) = 8.104, p = 0.000), angular depth (M = 351.2° vs. 437°; t(11)=4.527, p = 0.000), spiral diameter (M = 3.24 mm vs. 5.027 mm; t(11)=11.1, p = 0.000), and modified ICPI (M = 0.628 vs. 0.783; t(11)=9.742, p = 0.000). eABR results showed larger average amplitudes with straight arrays in basal and mid regions (p < 0.05), suggesting potentially enhanced neural stimulation and a more focused stimulation with perimodiolar arrays.

CONCLUSION

Our findings underscore the importance of electrode array type in CI outcomes, emphasizing the need for precise positioning to optimize neural stimulation and clinical outcomes. Personalized array selection based on CBCT assessments may lead to improved speech perception, pitch discrimination, and overall auditory function. Future research should aim to refine stimulation strategies to maximize CI benefits, particularly where direct measurements strongly correlate with clinical outcomes compared to derived measurements.

Для цитирования: Alnafjan Fadwa, da Cruz Melville J, McMahon Catherine M. Characteristics of Medial and Lateral Wall Cochlear Implant Arrays Demonstrated with Cone Beam CT and Evoked Electrical Auditory Brainstem Responses.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2025. DOI: 10.1097/mao.0000000000004431.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.