3D printed aiming device for equine proximal interphalangeal joint arthrodesis: Ex vivo study.

Aimonetti Emma, Genton Martin, Lischer Christoph Johannes, Rossignol Fabrice

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To determine the benefit of using a three-dimensional (3D)-printed titanium aiming device for placing two 5.5-mm transarticular screws during proximal interphalangeal joint (PIPJ) arthrodesis in equine species.

STUDY

DESIGN

Ex vivo study.

SAMPLE POPULATION: A total of 40 cadaveric forelimbs from adult horses.

METHODS

A senior surgeon and a first-year resident each performed 20 constructs: 10 using the device and 10 using a freehand technique. Procedure duration and number of intraoperative radiographs were recorded. Final radiographs were evaluated by an ECVS-certified surgeon unaware of treatment group, based on two criteria: compliance with the AO principles, and symmetry and homogeneity of the construct. Following implants removal, the position of the thread hole on the articular surface of the second phalanx was assessed and graded for accuracy.

RESULTS

For the inexperienced surgeon, the device reduced mean procedure time from 27.24 to 20.16 min (-25%, p < .01), decreased mean radiograph number from 6.1 to 2.1 (-60%, p < .01), and improved mean accuracy scores from 4.4 to 5.8 (+31%, p < .01). Construct symmetry also improved from 1.5 to 2.2 (+47%, p < .05). For the senior surgeon, only the number of radiographs was significantly reduced, from 3.1 to 1.3 (-58%, p < .01), with no significant difference in duration, accuracy, or symmetry.

CONCLUSION

The aiming device improved feasibility, accuracy, and repeatability of transarticular screw placement for the inexperienced surgeon.

IMPACT: This 3D-printed device appears to be as a promising tool for achieving gold-standard transarticular screw placement and may increases the accuracy of this surgery.

Для цитирования: Aimonetti Emma, Genton Martin, Lischer Christoph Johannes, Rossignol Fabrice. 3D printed aiming device for equine proximal interphalangeal joint arthrodesis: Ex vivo study.. Veterinary surgery : VS. 2026. DOI: 10.1111/vsu.70085.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.