Examination of the Construct Validity of Nextiles Arm Sleeve as a Measure of Peak Elbow-Varus Torque.

Wolf Jakob, Nicholson Kristen, Fernandez Garrett, Impellizzeri Franco M, Ward Patrick, Shrier Ian, Martin Chelsea, O'Connell Nathan, Thigpen Charles, Shanley Ellen, Bullock Garrett

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

CONTEXT: New wearable circuited fabric biomechanic technology that can be used in a more ecological context to measure elbow-varus torque and throw count is now available but needs to be validated compared with markerless biomechanics.

OBJECTIVE

To examine the construct validity of the Nextiles Arm Sleeve by assessing its agreement and convergence with markerless 3-dimensional (3D) biomechanics used as a reference for (1) peak elbow-varus torque and (2) pitches thrown with separate analyses for fastballs, breaking balls, and changeups.

DESIGN

Cross-sectional study.

SETTING

Laboratory.

PATIENTS OR OTHER

PARTICIPANTS

A total of 29 collegiate pitchers (age = 19.5 ± 1.3 years, height = 1.88 ± 0.06 m, mass = 91.7 ± 9.5 kg, body mass index = 26.2 ± 1.9 kg/m2) participated.

INTERVENTIONS: Pitchers were assessed simultaneously via the Nextiles sleeve and markerless 3D motion capture (KinaTrax) as they threw fastballs, breaking balls, and changeups to a catcher at regulation distance.

MAIN OUTCOME MEASURES: Nested Bland-Altman limits of agreement (LoA) for construct validity, mixed-effects linear regressions for convergence validity, and intraclass correlation coefficients (ICCs) for test-retest reliability were calculated between the Nextiles sleeve and 3D biomechanics for elbow-varus torque. We determined that an a priori acceptable threshold for LoA and convergence was a 95% CI width of 0.15 and 95% CI lower limit of 0.50.

RESULTS

Pitchers threw 200 pitches (fastball: 72%, n = 143; breaking ball: 14%, n = 27; changeup: 15%, n = 30), with 182 pitches recorded by the Nextiles sleeve and 18 pitches (n = 3 pitchers) not recorded due to technical errors. Agreement and validity for pitch measures (mean difference [lower, upper] LoA = 6.7 N·m [-26.2, 39.6 N·m]; β = 0.31 [0.12, 0.52]; Nextiles sleeve ICC [95% CI] = 0.96 [0.93, 0.98]; KinaTrax ICC [95% CI] = 0.97 [0.94, 0.98]) were higher compared with agreement for body weight × height (BW×H; mean difference [lower, upper] LoA = 0.37 BW×H [-1.41, 2.45 BW×H]; β = 0.37 [-1.41, 2.15]; Nextiles sleeve ICC [95% CI] = 0.93 [0.88, 0.96]; KinaTrax ICC [95% CI] = 0.95 [0.92, 0.98]). We found no interaction per pitch type (P = .47) or pitch thrown (P = .58).

CONCLUSIONS

The Nextiles sleeve collected pitch counts accurately except for 3 pitchers for whom it did not collect any pitches thrown due to technical errors. However, the demonstrated LoA and convergence were beyond the a priori acceptability threshold and lower than the a priori defined acceptable lower bound of the 95% CI. These results suggest that the Nextiles sleeve does not measure the same construct as KinaTrax.

Для цитирования: Wolf Jakob, Nicholson Kristen, Fernandez Garrett, Impellizzeri Franco M, Ward Patrick, Shrier Ian, Martin Chelsea, O'Connell Nathan, Thigpen Charles, Shanley Ellen, Bullock Garrett. Examination of the Construct Validity of Nextiles Arm Sleeve as a Measure of Peak Elbow-Varus Torque.. Journal of athletic training. 2026. DOI: 10.4085/1062-6050-0547.25.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.