MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s43390-024-00825-y

Медицинская библиотека · Spine deformity · 2024

Spine surgeon versus AI algorithm full-length radiographic measurements: a validation study of complex adult spinal deformity patients.

Haselhuhn Jason J, Soriano Paul Brian O, Grover Priyanka, Dreischarf Marcel, Odland Kari, Hendrickson Nathan R, Jones Kristen E, Martin Christopher T, Sembrano Jonathan N, Polly David W

СтатьяВалидационное исследованиеСравнительное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Spinal measurements play an integral role in surgical planning for a variety of spine procedures. Full-length imaging eliminates distortions that can occur with stitched images. However, these images take radiologists significantly longer to read than conventional radiographs. Artificial intelligence (AI) image analysis software that can make such measurements quickly and reliably would be advantageous to surgeons, radiologists, and the entire health system.

MATERIALS AND

METHODS

Institutional Review Board approval was obtained for this study. Preoperative full-length standing anterior-posterior and lateral radiographs of patients that were previously measured by fellowship-trained spine surgeons at our institution were obtained. The measurements included lumbar lordosis (LL), greatest coronal Cobb angle (GCC), pelvic incidence (PI), coronal balance (CB), and T1-pelvic angle (T1PA). Inter-rater intra-class correlation (ICC) values were calculated based on an overlapping sample of 10 patients measured by surgeons. Full-length standing radiographs of an additional 100 patients were provided for AI software training. The AI algorithm then measured the radiographs and ICC values were calculated.

RESULTS

ICC values for inter-rater reliability between surgeons were excellent and calculated to 0.97 for LL (95% CI 0.88-0.99), 0.78 (0.33-0.94) for GCC, 0.86 (0.55-0.96) for PI, 0.99 for CB (0.93-0.99), and 0.95 for T1PA (0.82-0.99). The algorithm computed the five selected parameters with ICC values between 0.70 and 0.94, indicating excellent reliability. Exemplary for the comparison of AI and surgeons, the ICC for LL was 0.88 (95% CI 0.83-0.92) and 0.93 for CB (0.90-0.95). GCC, PI, and T1PA could be determined with ICC values of 0.81 (0.69-0.87), 0.70 (0.60-0.78), and 0.94 (0.91-0.96) respectively.

CONCLUSIONS

The AI algorithm presented here demonstrates excellent reliability for most of the parameters and good reliability for PI, with ICC values corresponding to measurements conducted by experienced surgeons. In future, it may facilitate the analysis of large data sets and aid physicians in diagnostics, pre-operative planning, and post-operative quality control.

Для цитирования: Haselhuhn Jason J, Soriano Paul Brian O, Grover Priyanka, Dreischarf Marcel, Odland Kari, Hendrickson Nathan R, Jones Kristen E, Martin Christopher T, Sembrano Jonathan N, Polly David W. Spine surgeon versus AI algorithm full-length radiographic measurements: a validation study of complex adult spinal deformity patients.. Spine deformity. 2024. DOI: 10.1007/s43390-024-00825-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.