Intra and interrater reliability of spinal sagittal curves and mobility using pocket goniometer IncliMed® in healthy subjects.

Alderighi Marzia, Ferrari Raffaello, Maghini Irene, Del Felice Alessandra, Masiero Stefano

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Radiographic examination is the gold standard to evaluate spine curves, but ionising radiations limit routine use. Non-invasive methods, such as skin-surface goniometer (IncliMed®) should be used instead.

OBJECTIVE

To evaluate intra- and interrater reliability to assess sagittal curves and mobility of the spine with IncliMed®.

METHODS

a reliability study on agonistic football players. Thoracic kyphosis, lumbar lordosis and mobility of the spine were assessed by IncliMed®. Measurements were repeated twice by each examiner during the same session with between-rater blinding. Intrarater and interrater reliability were measured by Intraclass Correlation Coefficient (ICC), 95% Confidence Interval (CI 95%) and Standard Error of Measurement (SEM).

RESULTS

Thirty-four healthy female football players (19.17 ± 4.52 years) were enrolled. Statistical results showed high intrarater (0.805-0.923) and interrater (0.701-0.886) reliability (ICC > 0.8). The obtained intra- and interrater SEM were low, with overall absolute intrarater values between 1.39° and 2.76° and overall interrater values between 1.71° and 4.25°.

CONCLUSIONS

IncliMed® provides high intra- and interrater reliability in healthy subjects, with limited Standard Error of Measurement. These results encourage its use in clinical practice and scientific research.

Для цитирования: Alderighi Marzia, Ferrari Raffaello, Maghini Irene, Del Felice Alessandra, Masiero Stefano. Intra and interrater reliability of spinal sagittal curves and mobility using pocket goniometer IncliMed® in healthy subjects.. Journal of back and musculoskeletal rehabilitation. 2016. DOI: 10.3233/bmr-160704.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.