Subsecond and Submillimeter Resolution Positional Verification for Stereotactic Irradiation of Spinal Lesions.

Hazelaar Colien, Dahele Max, Mostafavi Hassan, van der Weide Lineke, Slotman Ben J, Verbakel Wilko F A R

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: Spine stereotactic body radiation therapy (SBRT) requires highly accurate positioning. We report our experience with markerless template matching and triangulation of kilovoltage images routinely acquired during spine SBRT, to determine spine position.

METHODS AND MATERIALS: Kilovoltage images, continuously acquired at 7, 11 or 15 frames/s during volumetric modulated spine SBRT of 18 patients, consisting of 93 fluoroscopy datasets (1 dataset/arc), were analyzed off-line. Four patients were immobilized in a head/neck mask, 14 had no immobilization. Two-dimensional (2D) templates were created for each gantry angle from planning computed tomography data and registered to prefiltered kilovoltage images to determine 2D shifts between actual and planned spine position. Registrations were considered valid if the normalized cross correlation score was ≥0.15. Multiple registrations were triangulated to determine 3D position. For each spine position dataset, average positional offset and standard deviation were calculated. To verify the accuracy and precision of the technique, mean positional offset and standard deviation for twenty stationary phantom datasets with different baseline shifts were measured.

RESULTS

For the phantom, average standard deviations were 0.18 mm for left-right (LR), 0.17 mm for superior-inferior (SI), and 0.23 mm for the anterior-posterior (AP) direction. Maximum difference in average detected and applied shift was 0.09 mm. For the 93 clinical datasets, the percentage of valid matched frames was, on average, 90.7% (range: 49.9-96.1%) per dataset. Average standard deviations for all datasets were 0.28, 0.19, and 0.28 mm for LR, SI, and AP, respectively. Spine position offsets were, on average, -0.05 (range: -1.58 to 2.18), -0.04 (range: -3.56 to 0.82), and -0.03 mm (range: -1.16 to 1.51), respectively. Average positional deviation was <1 mm in all directions in 92% of the arcs.

CONCLUSIONS

Template matching and triangulation using kilovoltage images acquired during irradiation allows spine position detection with submillimeter accuracy at subsecond intervals. Although the majority of patients were not immobilized, most vertebrae were stable at the sub-mm level during spine SBRT delivery.

Для цитирования: Hazelaar Colien, Dahele Max, Mostafavi Hassan, van der Weide Lineke, Slotman Ben J, Verbakel Wilko F A R. Subsecond and Submillimeter Resolution Positional Verification for Stereotactic Irradiation of Spinal Lesions.. International journal of radiation oncology, biology, physics. 2016. DOI: 10.1016/j.ijrobp.2016.01.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.