Utility of Intraoperative Radiographs in Pediatric Cochlear Implant Surgery.
Аннотация
OBJECTIVE
To evaluate the role of intraoperative radiographs to confirm electrode position following pediatric cochlear implantation (CI).
STUDY
DESIGN
Retrospective chart review.
SETTING
Single tertiary care pediatric center.
METHODS
A retrospective chart review was conducted, including all pediatric patients undergoing CI at UPMC Children's Hospital of Pittsburgh over a 13-year period.
RESULTS
We identified 326 patients undergoing 492 procedures. Across the cohort, there were 7 cases that required intraoperative electrode reinsertion due to malposition or presumed malposition. For 6 of the 7 cases, intraoperative X-ray identified electrode malposition. Neural response telemetry (NRT) testing was also abnormal for 4 of these cases prior to reinsertion. Implantation of Cochlear's Slim Modiolar electrode was associated with an abnormal perioperative X-ray (odds ratio [OR]: 9.2, p = 0.03) and increased change in management (OR: 9.2, p = 0.03) compared to Cochlear's Contour Advance (CA). Incidence of abnormal X-rays was 1.24% overall, 4% in the Slim Modiolar group, and 0.3% in the CA group. The Slim Modiolar electrode accounted for 4 of 7 cases requiring reinsertion, and in all 4 of these cases, electrode fold-over was identified on the X-ray. NRT was normal in 1 of these 4 cases.
CONCLUSION
The use of Cochlear's Slim Modiolar electrode was associated with a significantly increased risk of abnormal intraoperative X-ray compared to the CA electrode. Given the risk of fold-over with routine insertion and normal electrical testing using the Slim Modiolar electrode, we recommend routine use of intraoperative skull X-ray to confirm electrode position.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.