Scandinavian journal of gastroenterology · 2000
Ultrasound-guided biopsies of abdominal organs with an automatic biopsy system. A retrospective analysis of the quality of biopsies and of hemorrhagic complications.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Riemann B, Menzel J, Schiemann U, Domschke W, Konturek J W
Аннотация
BACKGROUND: Ultrasound-guided biopsies of abdominal organs are not without risks for the patients; in particular, hemorrhagic complications may occur. Thus, over the last few years, automatic biopsy guns have been developed to facilitate the biopsy process.
METHODS: The aim of our retrospective study was to examine the quality of specimens and the complication rate of ultrasound-guided biopsies of abdominal organs carried out in our institution using the automatic Autovac biopsy system during a period of 1.5 years. Of the total number of 321 biopsies, 290 were performed with the 1.2-mm Autovac needle, and in 31 cases the 0.95-mm needle was used. Among the 321 biopsies there were 211 of the liver parenchyma (66%), 47 of a liver tumor (14%), 38 of the pancreas (12%), 15 of the kidney parenchyma (5%), and 10 of a retroperitoneal tumor (3%).
RESULTS: In 310 of the 321 biopsies it was possible to obtain sufficient diagnostically usable material for the pathologist (96.6%). In the other 11 cases the material obtained did not enable proper histologic diagnosis (3.4%). Two of these 11 biopsies were carried out with the 0.95-mm needle, and the other 9 with the 1.2-mm needle. Twenty-four hours after the biopsy each patient underwent routine ultrasound examination to exclude a possible bleeding. In eight cases an afterbleeding occurred (total hemorrhagic rate, 2.5%), four times without clinical consequences. The other four bleeding complications were more serious (1.2% of all taps), and all occurred after liver biopsies in patients with a history of liver complaints and abnormal clotting variables. There were no fatalities among our biopsies (mortality rate, 0%).
CONCLUSION: The automatic Autovac biopsy system is suitable and relatively safe for obtaining sufficient histopathologic material from intra-abdominal organs.
METHODS: The aim of our retrospective study was to examine the quality of specimens and the complication rate of ultrasound-guided biopsies of abdominal organs carried out in our institution using the automatic Autovac biopsy system during a period of 1.5 years. Of the total number of 321 biopsies, 290 were performed with the 1.2-mm Autovac needle, and in 31 cases the 0.95-mm needle was used. Among the 321 biopsies there were 211 of the liver parenchyma (66%), 47 of a liver tumor (14%), 38 of the pancreas (12%), 15 of the kidney parenchyma (5%), and 10 of a retroperitoneal tumor (3%).
RESULTS: In 310 of the 321 biopsies it was possible to obtain sufficient diagnostically usable material for the pathologist (96.6%). In the other 11 cases the material obtained did not enable proper histologic diagnosis (3.4%). Two of these 11 biopsies were carried out with the 0.95-mm needle, and the other 9 with the 1.2-mm needle. Twenty-four hours after the biopsy each patient underwent routine ultrasound examination to exclude a possible bleeding. In eight cases an afterbleeding occurred (total hemorrhagic rate, 2.5%), four times without clinical consequences. The other four bleeding complications were more serious (1.2% of all taps), and all occurred after liver biopsies in patients with a history of liver complaints and abnormal clotting variables. There were no fatalities among our biopsies (mortality rate, 0%).
CONCLUSION: The automatic Autovac biopsy system is suitable and relatively safe for obtaining sufficient histopathologic material from intra-abdominal organs.