Long-term outcome of transvenous bipolar atrial leads implanted in children and young adults with congenital heart disease.

Silvetti Massimo Stefano, Drago Fabrizio, Ravà Lucilla

СтатьяПеревод в очереди

Аннотация

AIMS: Atrial leads are often implanted in paediatric patients needing a pacemaker (PM). The aim of this study is the evaluation of their outcome in young patients.

METHODS AND

RESULTS

We evaluated transvenous atrial leads outcome in children and young adults from a single centre, with a retrospective analysis. A P< 0.05 was considered significant. Between 1992 and 2008, 110 patients, 75 with congenital heart defects (d-transposition of great arteries status/post, s/p, Mustard 41%, atrioventricular septal defect 11%, tetralogy 9%, ventricular septal defect 8%), aged 13.3 ± 5.3 years, underwent PM implantation with bipolar atrial transvenous leads for sinus node dysfunction (50%), atrioventricular block (38%), cardiomyopathies, and primary ventricular arrhythmias (12%). Leads are steroid-eluting (98%), tined (59%), screw-in (41%), polyurethane-insulated (72%), silicone-insulated (28%), and have been positioned by transcutaneous puncture of subclavian vein into right atrial appendage/remnant (RAA, 50%), right atrial free wall/septum (25%), left atrium (s/p Mustard, 25%). Follow-up duration is 6.4 ± 4.8 (range 0.1-18) years. At multivariate analysis, younger age at implant was a risk factor for lead failure (4 leads, 3.5%) (P= 0.03); 16 leads (14%) dislodged post-implantation and 12 were successfully repositioned, the others extracted or abandoned. Dislocation occurred more frequently with screw-in leads (P= 0.03) positioned outside RAA (P= 0.02). Atrial threshold showed a small but significant increase, 0.002 V/month (P< 0.001), impedance showed a decrease (0.6 Ω/month, P< 0.001), P-wave showed no significant difference.

CONCLUSIONS

Transvenous bipolar atrial leads have good long-term results in young patients, with a very low rate of lead failure. Older age at implant can further reduce this rate. Lead dislodgement is frequent in the post-operative period.

Для цитирования: Silvetti Massimo Stefano, Drago Fabrizio, Ravà Lucilla. Long-term outcome of transvenous bipolar atrial leads implanted in children and young adults with congenital heart disease.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2012. DOI: 10.1093/europace/eus024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.