Antes Sebastian, Salah Mohamed, Linsler Stefan, Tschan Christoph A, Breuskin David, Oertel Joachim
Аннотация
INTRODUCTION: Aqueductoplasty as well as aqueductal stenting is an accepted therapy option in short-segment aqueductal stenosis and isolated fourth ventricle. Over the years, different techniques with only slight modifications by using a conventional neuroendoscope with a working sheath to introduce different instruments have been presented. In summary, the use of Fogarty balloon catheters or flexible endoscopes to pass the narrowed aqueduct is recommended.
METHODS: This technical report describes a substantially new technique for this purpose. Six patients underwent aqueductal stenting with a new intracatheter endoscope.
RESULTS: Aqueductal stenting was possible in 4 out of 6 cases. No complications occurred. Handling of this new technique was good and easy without a prolonged learning curve. All four stents did work appropriately, and the procedure was considered to be successful. Of the two failures, the technique was abandoned and endoscopic third ventriculostomy (ETV) was performed in one. In the other case, suboccipital shunting was done.
CONCLUSION: This technical report describes a substantially new technique for aqueductal stenting. The combination of an intracatheter miniature endoscope and a prepared ventricular catheter enables careful and elegant aqueductal stenting. Large or flexible endoscopes, balloons, or special instruments to place a stent have become completely obsolete in selected cases.
METHODS: This technical report describes a substantially new technique for this purpose. Six patients underwent aqueductal stenting with a new intracatheter endoscope.
RESULTS: Aqueductal stenting was possible in 4 out of 6 cases. No complications occurred. Handling of this new technique was good and easy without a prolonged learning curve. All four stents did work appropriately, and the procedure was considered to be successful. Of the two failures, the technique was abandoned and endoscopic third ventriculostomy (ETV) was performed in one. In the other case, suboccipital shunting was done.
CONCLUSION: This technical report describes a substantially new technique for aqueductal stenting. The combination of an intracatheter miniature endoscope and a prepared ventricular catheter enables careful and elegant aqueductal stenting. Large or flexible endoscopes, balloons, or special instruments to place a stent have become completely obsolete in selected cases.