Serum bile acids profiling by liquid chromatography-tandem mass spectrometry (LC-MS/MS) and its application on pediatric liver and intestinal diseases.

Fu Xiaowei, Xiao Yi, Golden Jamie, Niu Sizhe, Gayer Christopher P

СтатьяПеревод в очереди

Аннотация

Background A method for bile acid profiling measuring 21 primary and secondary bile acids in serum samples was developed and validated with liquid chromatography-tandem mass spectrometry (LC-MS/MS). Sample preparation included spiking with internal standards followed by protein precipitation, centrifugation, drying under nitrogen gas and reconstitution. Extracted samples were injected onto a Phenomenex Kinetex C18 column (150 × 4.60 mm, 2.6 μm). Methods Data was collected with LC-MS/MS operated in negative ion mode with multiple reaction monitoring (MRM) and single reaction monitoring (SRM). The analytical run time was 12 min. Results The method showed excellent linearity with high regression coefficients (>0.99) over a range of 0.05 and 25 μM for all analytes tested. The method also showed acceptable intra-day and inter-day accuracy and precision. As a proof of concept, the analytical method was applied to patients with neonatal intrahepatic cholestasis caused by citrin deficiency (NICCD), biliary atresia (BA), and necrotizing enterocolitis (NEC), and distinct bile acids profiles were demonstrated. Conclusions The method could be poised to identify possible biomarkers for non-invasive early diagnosis of these disorders.

Для цитирования: Fu Xiaowei, Xiao Yi, Golden Jamie, Niu Sizhe, Gayer Christopher P. Serum bile acids profiling by liquid chromatography-tandem mass spectrometry (LC-MS/MS) and its application on pediatric liver and intestinal diseases.. Clinical chemistry and laboratory medicine. 2020. DOI: 10.1515/cclm-2019-0354.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.