MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1743-6109.2012.02910.x

Медицинская библиотека · The journal of sexual medicine · 2012

A new device for simultaneous measurement of pelvic floor muscle activity and vaginal blood flow: a test in a nonclinical sample.

Both Stephanie, van Lunsen Rik, Weijenborg Philomeen, Laan Ellen

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Dyspareunia in women, defined as persistent or recurrent genital pain associated with sexual intercourse, is hypothesized to be related to (fear associated) pelvic floor hyperactivity and to diminished sexual arousal. Psychophysiological research to support these hypotheses is scarce and concentrates mostly on the role of either pelvic floor activity or sexual arousal. To investigate both factors, a measurement device that enables simultaneous assessment of pelvic floor muscle activity and genital sexual arousal would be most optimal.

AIM: The aim of this study was to test a new vaginal device0-a vaginal photoplethysmograph with build-in surface electromyography (EMG)--that allows simultaneous assessment of pelvic floor muscle activity and vaginal blood flow.

MAIN OUTCOME MEASURES: Genital arousal measured as vaginal pulse amplitude (VPA) and vaginal surface EMG.

METHODS

Thirty-six sexually functional women participated. To investigate the sensitivity of the device for changes in genital blood flow and involuntary changes in pelvic floor activity, VPA and vaginal surface EMG were monitored during exposure to sexual and anxiety-evoking film clips. In addition, vaginal surface EMG was monitored during voluntary flick and hold contractions.

RESULTS

VPA increased in response to the sexual film, and EMG values were significantly higher in response to the anxiety-evoking film. Higher EMG values in response to the anxiety film were associated with lower VPA. EMG during the instructed 3-second hold pelvic floor contractions showed, as expected, higher values during pelvic floor contractions with support of surrounding muscle groups, compared with pelvic floor muscles alone.

CONCLUSION

The device is sensitive to changes in vaginal blood flow in response to sexual stimuli, and it is able to pick up small, involuntary changes in pelvic floor activity associated with anxiety. Also, the device is able to record changes in pelvic floor activity during voluntary pelvic floor contractions. This new device will be a valuable tool in further research on superficial dyspareunia.

Для цитирования: Both Stephanie, van Lunsen Rik, Weijenborg Philomeen, Laan Ellen. A new device for simultaneous measurement of pelvic floor muscle activity and vaginal blood flow: a test in a nonclinical sample.. The journal of sexual medicine. 2012. DOI: 10.1111/j.1743-6109.2012.02910.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.