Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.

Zhou Yang, Liu Yimei, Chen Meining, Fang Jianlan, Xiao Liangjie, Huang Shaomin, Qi Zhenyu, Deng Xiaowu, Zhang Jun, Peng Yinglin

СтатьяПеревод в очереди

Аннотация

PURPOSE: We aimed to perform the commissioning and clinical evaluation of myQA SRS detector array for patient-specific quality assurance (PSQA) of stereotactic radiosurgery (SRS)/ stereotactic body radiotherapy (SBRT) plans.

METHODS

To perform the commissioning of myQA SRS, its dose linearity, dose-rate dependence, angular dependence, and field-size dependence were investigated. Ten SBRT plans were selected for clinical evaluation: 1) Common clinical deviations based on the original SBRT plan (Plan0), including multileaf collimator (MLC) positioning deviation and treatment positioning deviation were introduced. 2) Compared the performance of the myQA SRS and a high-resolution EPID dosimetry system in PSQA measurement for the SBRT plans. Evaluation parameters include gamma passing rate (GPR) and distance-to-agreement (DTA) pass rate (DPR).

RESULTS

The dose linearity, angle dependence, and field-size dependence of myQA SRS system exhibit excellent performance. The myQA SRS is highly sensitive in the detection of MLC deviations. The GPR of (3%/1 mm) decreases from 90.4% of the original plan to 72.7%/62.9% with an MLC outward/inward deviation of 3 mm. Additionally, when the setup error deviates by 1 mm in the X, Y, and Z directions with the GPR of (3%/1 mm) decreasing by an average of -20.9%, -25.7%, and -24.7%, respectively, and DPR (1 mm) decreasing by an average of -33.7%, -32.9%, and -29.8%. Additionally, the myQA SRS has a slightly higher GPR than EPID for PSQA, However, the difference is not statistically significant with the GPR of (3%/1 mm) of (average 90.4%% vs. 90.1%, p = 0.414).

CONCLUSION

Dosimetry characteristics of the myQA SRS device meets the accuracy and sensitivity requirement of PSQA for SRS/SBRT treatment. The dose rate dependence should be adequately calibrated before its application and a more stringent GPR (3%/1 mm) evaluation criterion is suggested when it is used for SRS/SBRT QA.

Для цитирования: Zhou Yang, Liu Yimei, Chen Meining, Fang Jianlan, Xiao Liangjie, Huang Shaomin, Qi Zhenyu, Deng Xiaowu, Zhang Jun, Peng Yinglin. Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.. Journal of applied clinical medical physics. 2024. DOI: 10.1002/acm2.14258.