MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00586-016-4786-8

Медицинская библиотека · European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2016

Rod fracture and lengthening intervals in traditional growing rods: is there a relationship?

Hosseini Pooria, Pawelek Jeff B, Nguyen Stacie, Thompson George H, Shah Suken A, Flynn John M, Dormans John P, Akbarnia Behrooz A, Group Growing Spine Study

СтатьяПеревод в очереди

Аннотация

PURPOSE: Is there any relationship between lengthening intervals and rod fracture in traditional growing rod (TGR) surgery?

METHODS

A multicenter EOS database was queried for patients who had: (1) dual growing rods for EOS; (2) minimum 2-year follow-up; (3) a minimum of 2 lengthenings; and (4) revision surgery due to rod fracture. Of 138 patients who met the criteria: 56 patients experienced at least one-rod fracture (RF group) and 82 patients had no rod fractures (NRF group). In addition to each patient's lengthening intervals, demographics, construct details, and radiographic parameters were compared.

RESULTS

RF and NRF patients had a mean pre-op age of 5.7 years (range 1.3-10.7) and 7.3 years (range 1.6-12.8), respectively (p < 0.001). There was no significant association between etiologies and rod fracture or between BMI and rod fracture (p = 0.979). There was no significant difference between lengthening intervals between the RF and NRF groups (p > 0.05). RF and NRF patients had statistically similar mean pre-op major curve size and max kyphosis (p = 0.279; p = 0.619, respectively). Stainless steel rods fractured more frequently compared with Titanium rods (SS 49.2 % vs. Ti 38 %; p = 0.004). Rod fracture occurred more in rods smaller than 4 mm (p = 0.011).

CONCLUSIONS

Lengthening intervals were not statistically different in RF and NRF groups and there was no association between lengthening interval and rod fracture in TGR cases. It was shown that patients who had rod fracture were younger and were more likely to have SS rods with smaller than 4 mm diameter.

Для цитирования: Hosseini Pooria, Pawelek Jeff B, Nguyen Stacie, Thompson George H, Shah Suken A, Flynn John M, Dormans John P, Akbarnia Behrooz A, Group Growing Spine Study. Rod fracture and lengthening intervals in traditional growing rods: is there a relationship?. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2016. DOI: 10.1007/s00586-016-4786-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.