DOI 10.1111/pace.14018

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2020

Safety of bioelectrical impedance analysis in advanced heart failure patients.

Roehrich Luise, Suendermann Simon, Just Isabell Anna, Knierim Jan, Mulzer Johanna, Mueller Marcus, Eulert-Grehn Jaime-Juergen, Hummel Manfred, Starck Christoph, Potapov Evgenij, Knosalla Christoph, Falk Volkmar, Schoenrath Felix

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Cardiac cachexia and frailty are major complications of advanced heart failure (AHF). Bioelectrical impedance analysis (BIA) may provide valuable information regarding fluid balance, muscle mass and prognosis. The main concerns regarding the use of BIA in AHF patients remain arrhythmias and electromagnetic interferences with cardiac implantable electronic devices (CIEDs). Reliable data regarding patients on continuous-flow ventricular assist device (cf-VAD) remain scarce. The aim of this study is to evaluate the safety of BIA in AHF patients on pro-arrhythmogenic therapy with an implanted CIED and/or with a cf-VAD.

METHODS

We prospectively performed 217 BIA measurements in 143 AHF patients at risk of severe arrhythmias due to inotropic support/a history of ventricular arrhythmias and/or treated with CIED, including 104 patients with an ICD, CRT or pacemaker and 95 patients with a cf-VAD. All patients were under continuous Electrocardiogram (ECG) monitoring and clinical surveillance for 24 hours.

RESULTS

No adverse events were observed during the 217 BIA measurements: No rhythm disturbances were documented in the telemetric monitoring during or within 30 minutes after the measurement. CIEDs showed no malfunction, regardless of the location measured or the device manufacturer. In particular, no inappropriate shocks were observed. No alarms, flow disturbances, or malfunctions of the cf-VAD occurred during or after the measurements.

CONCLUSION

We consider BIA a safe measurement with major clinical relevance in our cohort of AHF patients, despite an increased arrhythmic potential on inotropic support or the presence of implanted electronic devices (ICD, CRT, pacemaker and cf-VAD).

Для цитирования: Roehrich Luise, Suendermann Simon, Just Isabell Anna, Knierim Jan, Mulzer Johanna, Mueller Marcus, Eulert-Grehn Jaime-Juergen, Hummel Manfred, Starck Christoph, Potapov Evgenij, Knosalla Christoph, Falk Volkmar, Schoenrath Felix. Safety of bioelectrical impedance analysis in advanced heart failure patients.. Pacing and clinical electrophysiology : PACE. 2020. DOI: 10.1111/pace.14018.

Предметные рубрики MeSH