BACKGROUND

Stereophotogrammetry (SPG) provides a more objective measurement of burn wound area than traditional clinical assessments. The recently developed Intel® RealSense™ D415/Wound Measure SPG system has yet to undergo formal evaluation in a paediatric population.

METHODS

A pilot study comparing the Intel® RealSense™ D415/Wound Measure to the previously validated LifeVizII®/DermaPix® SPG system, for burn assessment was conducted at a tertiary paediatric burn centre. Both systems were used to photograph and analyse the same acute wounds for each patient. Three independent raters measured wound area and perimeter. Level of agreement between systems and raters was estimated by calculating the intra-class correlation coefficient.

RESULTS

Wound area measurements were completed in both systems for 25 burns from 13 patients (median age, 2 years). The participants were mainly female (n = 9), with a median TBSA-B of 9% (IQR 3-20%). There was strong agreement between the systems 0.757 (95% CI 0.521, 0.885, p < 0.001). Within each SPG system, there was excellent inter-rater reliability.

CONCLUSION

The Intel® RealSense™ D415/Wound Measure system may be a viable addition to the clinician's toolkit in the assessment of paediatric burn wound area. As with other SPG systems, there were significant challenges measuring wounds to highly contoured surfaces.

Для цитирования: Bairagi Anjana, Frear Cody, Sweeting Caitlin, Kimble Roy, Griffin Bronwyn. A pilot study comparing two burn wound stereophotogrammetry systems in a paediatric population.. Burns : journal of the International Society for Burn Injuries. 2021. DOI: 10.1016/j.burns.2021.04.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.