Implantable direct-current bone stimulators in high-risk and revision foot and ankle surgery: a retrospective analysis with outcome assessment.
Аннотация
Efficacy and morbidity of a surgically implanted direct-current bone stimulator were evaluated in 38 patients (40 feet) with fracture nonunion or at high risk for nonunion; 14 of these patients had Charcot (diabetic) neuroarthropathy. Union occurred in 26 (65%) of the 40 feet; complications other than nonunion occurred in 16 feet (40%). Two amputations (5%) were performed in cases of intractable neuritis and deep infection. Of the 6 cases of deep infection (15%), 5 resolved with device removal, and the sixth case required below-knee amputation. Use of a bone stimulator in patients with diabetes may be problematic, but the device did not have any adverse effects in other high-risk patients.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.