MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1088/0031-9155/58/21/7715

Медицинская библиотека · Physics in medicine and biology · 2013

High-sensitivity brain SPECT system using cadmium telluride (CdTe) semiconductor detector and 4-pixel matched collimator.

Suzuki Atsuro, Takeuchi Wataru, Ishitsu Takafumi, Tsuchiya Katsutoshi, Morimoto Yuichi, Ueno Yuichiro, Kobashi Keiji, Kubo Naoki, Shiga Tohru, Tamaki Nagara

СтатьяПеревод в очереди

Аннотация

For high-sensitivity brain imaging, we have developed a two-head single-photon emission computed tomography (SPECT) system using a CdTe semiconductor detector and 4-pixel matched collimator (4-PMC). The term, '4-PMC' indicates that the collimator hole size is matched to a 2 × 2 array of detector pixels. By contrast, a 1-pixel matched collimator (1-PMC) is defined as a collimator whose hole size is matched to one detector pixel. The performance of the higher-sensitivity 4-PMC was experimentally compared with that of the 1-PMC. The sensitivities of the 1-PMC and 4-PMC were 70 cps/MBq/head and 220 cps/MBq/head, respectively. The SPECT system using the 4-PMC provides superior image resolution in cold and hot rods phantom with the same activity and scan time to that of the 1-PMC. In addition, with half the usual scan time the 4-PMC provides comparable image quality to that of the 1-PMC. Furthermore, (99m)Tc-ECD brain perfusion images of healthy volunteers obtained using the 4-PMC demonstrated acceptable image quality for clinical diagnosis. In conclusion, our CdTe SPECT system equipped with the higher-sensitivity 4-PMC can provide better spatial resolution than the 1-PMC either in half the imaging time with the same administered activity, or alternatively, in the same imaging time with half the activity.

Для цитирования: Suzuki Atsuro, Takeuchi Wataru, Ishitsu Takafumi, Tsuchiya Katsutoshi, Morimoto Yuichi, Ueno Yuichiro, Kobashi Keiji, Kubo Naoki, Shiga Tohru, Tamaki Nagara. High-sensitivity brain SPECT system using cadmium telluride (CdTe) semiconductor detector and 4-pixel matched collimator.. Physics in medicine and biology. 2013. DOI: 10.1088/0031-9155/58/21/7715.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.