Analysis of dose distribution around a computed tomography scanner in terms of exposure to scattered ionizing radiation of caregivers of pediatric patients.

Biegała Michał, Brodecki Marcin, Skoczylas Krystian, Jakubowska Teresa, Domienik-Andrzejewska Joanna

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

During computed tomography (CT), a large amount of ionizing radiation is emitted to ensure high quality of the obtained radiological image. This study measured the dose distribution around the CT scanner and the exposure of people staying near the CT scanner during the examination.

MATERIAL AND

METHODS

The measurements used an anthropomorphic phantom to assess human exposure to ionizing radiation. The probability of inducing leukemia and other cancers as a result of absorbing doses recorded around the CT device was also calculated.

RESULTS

The highest exposure to scattered radiation in the proximity of the CT scanner is recorded at the gantry of the tomograph, i.e., 55.7 μGy, and the lowest, below lower detection limit of 6 μGy at the end of the diagnostic table. The whole-body detector placed on the anthropomorphic phantom located at the diagnostic table right next to the CT gantry recorded 59.5 μSv and at the end of the table 1.5 μSv. The average doses to the lenses in these locations were: 32.1 μSv and 2.9 μSv, respectively.

CONCLUSIONS

The probability of induction of leukemia or other types of cancer is low, but the need for people to stay in the examination room during a CT examination should be limited to the necessary minimum. Int J Occup Med Environ Health. 2024;37(3):326-34.

Для цитирования: Biegała Michał, Brodecki Marcin, Skoczylas Krystian, Jakubowska Teresa, Domienik-Andrzejewska Joanna. Analysis of dose distribution around a computed tomography scanner in terms of exposure to scattered ionizing radiation of caregivers of pediatric patients.. International journal of occupational medicine and environmental health. 2024. DOI: 10.13075/ijomeh.1896.02386.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.