Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam.

Che Fru Leonard, Insley Benjamin A, Lee Dongyeon, Bartkoski Dirk A, Reiazi Reza, Schori Aviad, Tailor Ramesh, Salehpour Mohammad

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Objective. This work aimed to evaluate the dosimetric characteristics of the new HyperScintTMResearch Platform 200 system (HS-RP200), with orthovoltage beams and a novel kilovoltage converging beam, converging radiotherapy and radiosurgery (CRnR) beam.Approach. Several dosimetric properties of the HS-RP200 detector with several orthovoltage beams were studied in air. Relative dosimetry of the HS-RP200 detector was also studied for use on the converging x-ray beam. Energy correction factors were compared between measurement, Monte Carlo, and analytical methods.Main results.The signal-to-noise ratio for the detector increased as the frame exposure time increased, and repeated measurements had less uncertainty when measurements were performed without stem-effect removal (WOSER). The short-term stability study of the detector's signal showed an acceptable variance of its readings to within 1% during 1 h measurements. The detector exhibited angular independence within the measurement uncertainty of 1%. The detector exhibited excellent dose linearity response with the orthovoltage beams. During an 80 d period, repeated measurements determined that the detector possesses long-term stability within a root mean square (rms) of 2.3% and 0.6% when measurements were performed with stem effect removal and WOSER, respectively. Calculated energy correction factors implied significant fluctuations of the HS-RP200 detector with the beam quality of orthovoltage beams, ranging from 3% to 66%, depending on the divergence of the effective energy of the beam from reference quality. Monte Carlo-simulated energy correction factors agreed with measurements with a maximum difference of 5.4% when measurements were performed WOSER. Lastly, the HS-RP200 detector system exhibited good relative dosimetric characteristics with the CRnR beam. A maximum rms value of 0.019 was calculated for the normalized profiles when measurements from the HS-RP200 detector were compared to measurements from film.Significance. The HS-RP200 detector could be used as an alternative to other detectors with orthovoltage beams, including the CRnR beam.

Для цитирования: Che Fru Leonard, Insley Benjamin A, Lee Dongyeon, Bartkoski Dirk A, Reiazi Reza, Schori Aviad, Tailor Ramesh, Salehpour Mohammad. Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam.. Physics in medicine and biology. 2025. DOI: 10.1088/1361-6560/ae0efb.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.