Wearing an ultrasound probe during walking does not influence lower limb joint kinematics in adolescents with cerebral palsy and typically developing peers.

Cenni Francesco, Alexander Nathalie, Laatikainen-Raussi Iida, Sukanen Maria, Finni Taija

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Enhancing traditional three-dimensional gait analysis with a portable ultrasound device at the lower-limb muscle-tendon level enables direct measurement of muscle and tendon lengths during walking. However, it is important to consider that the size of the ultrasound probe and its attachment on the lower limb may potentially influence gait pattern.

RESEARCH QUESTION: What is the effect of wearing an ultrasound probe at the lower limb in adolescents with cerebral palsy and typically developing peers?

METHODS

Eleven individuals with cerebral palsy and nine age-matched typically developing peers walking barefoot at their self-selected speed were analyzed. Data collection occurred under three conditions: the reference condition (GAIT), and two conditions involving placement of the ultrasound probe over the distal medial gastrocnemius-Achilles tendon junction (MTJ) and over the medial gastrocnemius mid-belly to capture fascicles (FAS). Data processing included calculating differences between conditions using root mean square error (RMSE) for joint kinematics and comparing them to the overall mean difference. Additionally, Spearman correlations were calculated to examine the relationship between kinematic RMSEs and walking speed.

RESULTS

No significant differences in stance phase duration or walking speed were observed among the three conditions. Average RMSEs were below 5° for all parameters and condition comparisons in both groups. In both the TD and CP groups, RMSE values during the swing phase were higher than those during the stance phase for all joints. No significant correlations were found between height or body mass and swing phase RMSEs. In the CP group, there was a significant correlation between joint kinematics RMSEs and differences in walking speed at the hip, knee and ankle joints when comparing the MTJ condition with the GAIT condition.

SIGNIFICANCE: This study confirms joint kinematics alterations are smaller than 5° due to wearing to the leg an ultrasound probe during walking.

Для цитирования: Cenni Francesco, Alexander Nathalie, Laatikainen-Raussi Iida, Sukanen Maria, Finni Taija. Wearing an ultrasound probe during walking does not influence lower limb joint kinematics in adolescents with cerebral palsy and typically developing peers.. Gait & posture. 2024. DOI: 10.1016/j.gaitpost.2024.05.017.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.