MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.wneu.2022.04.120

Медицинская библиотека · World neurosurgery · 2022

Self-Shielding Gyroscopic Radiosurgery-A First Clinical Experience, Case Series, and Dosimetric Comparison.

Muacevic Alexander, Eder Michael Martin, Hofmann Theresa, Fürweger Christoph, Ehret Felix

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Self-shielding gyroscopic radiosurgery (GRS) represents a technical innovation in the field of stereotactic radiosurgery. GRS does not require a radiation vault and is optimized for radiosurgical treatments. Reports on its usage are limited. We describe the first clinical experience of GRS at our institution to assess the application of GRS in the treatment of cranial tumors. Moreover, we perform a dosimetric comparison to robotic radiosurgery (RRS) with vestibular schwannoma (VS) GRS patients.

METHODS

Patients who were treated with GRS between July and November 2021 were included. Patient, tumor, and dosimetric characteristics were retrospectively summarized and analyzed.

RESULTS

Forty-one patients with 48 intracranial tumors were included. Tumor entities mostly comprised VS, brain metastases, and meningiomas. The median prescription dose and isodose line were 13.5 Gy and 50.0% for benign neoplasia versus 20 Gy and 60.0% for malignant tumors, respectively. The mean planning target volume was 1.5 cubic centimeters. All patients received a single-fraction treatment without encountering any technical setup difficulties. Treatment plan comparisons with RRS revealed comparable plan characteristics, dose gradients, and organs at risk doses. Significant differences were detected concerning the new conformity index and number of monitor units per treatment (both P < 0.01).

CONCLUSIONS

This case series provides more evidence on the usage of self-shielding GRS in the management of cranial tumors. Dosimetric comparisons for VS cases revealed mostly equivalent dosimetric characteristics to RRS. Further clinical and physical analyses for GRS are underway.

Для цитирования: Muacevic Alexander, Eder Michael Martin, Hofmann Theresa, Fürweger Christoph, Ehret Felix. Self-Shielding Gyroscopic Radiosurgery-A First Clinical Experience, Case Series, and Dosimetric Comparison.. World neurosurgery. 2022. DOI: 10.1016/j.wneu.2022.04.120.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.