DOI 10.1016/j.prro.2018.05.006
Assessing Functionality and Benefits of Comprehensive Dose Volume Prescriptions: An International, Multi-Institutional, Treatment Planning Study in Spine Stereotactic Body Radiation Therapy.
Аннотация
PURPOSE: This study aimed to assess the effectiveness of multiple dose-volume specifications in minimizing interinstitutional, target-prescribed, dose variations for spine stereotactic body radiation therapy (SBRT).
METHODS AND MATERIALS: Seven institutions with a total of 10 treatment apparatuses participated in this study. SBRT plans for 3 representative spinal metastases were generated using 2 different protocols (Protocols 1 and 2) for target dose. While using just 2 target dose objectives (doses delivered to 95% and maximum point dose) in Protocol 1, 3 target dose constraints (doses delivered to 95% and 50% and maximum point dose) were defined in Protocol 2 with the intent to decrease target dose variation. A dose-volume histogram analysis was performed for the evaluated planning target volume (PTVevl) and critical neural structures such as the spinal cord and cauda equina.
RESULTS
Doses to the organs at risk were all maintained at the maximal tolerance in both protocols; however, the interinstitutional variation of the PTVevl dose-volume histograms was significantly decreased with Protocol 2. Furthermore, the mean PTVevl covered by the prescription dose was increased from 73.0% in Protocol 1 to 85.8% in Protocol 2. There were no differences in the mean values of the nearly maximum dose of the critical neural structures between 2 protocols.
CONCLUSIONS
In spine SBRT with the emphasis on preservation of critical neural structures, the target prescribed dose should be defined by using multiple dose-volume objectives to minimize user and apparatus-dependent dose variabilities for the spinal metastases that are adjacent to the critical neural structures.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.