Comparison of four techniques for spine stereotactic body radiotherapy: Dosimetric and efficiency analysis.

Aljabab Saif, Vellayappan Balamurugan, Vandervoort Eric, Bahm Jamie, Zohr Robert, Sinclair John, Caudrelier Jean-Michel, Szanto Janos, Malone Shawn

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The aim of this study is to compare the dosimetric differences between four techniques for spine stereotactic body radiotherapy (SBRT): CyberKnife (CK), volumetric modulated arc therapy (VMAT), and helical tomotherapy (HT) with dynamic jaws (HT-D) and fixed jaws (HT-F).

MATERIALS/

METHODS

Data from 10 patients were utilized. All patients were planned for 24 Gy in two fractions, with the primary objectives being: (a) restricting the maximum dose to the cord to ≤ 17 Gy and/or cauda equina to ≤ 20 Gy, and (b) to maximize the clinical target volume (CTV) to receive the prescribed dose. Treatment plans were generated by separate dosimetrists and then compared using velocity AI. Parameters of comparison include target volume coverage, conformity index (CI), gradient index (GI), homogeneity index (HI), treatment time (TT) per fraction, and monitor units (MU) per fraction.

RESULTS

PTV D2 and D5 were significantly higher for CK compared to VMAT, HT-F, and HT-D (P < 0.001). The average volume of CTV receiving the prescription dose (CTV D95) was significantly less for VMAT compared to CK, HT-F and HT-D (P = 0.036). CI improved for CK (0.69), HT-F (0.66), and HT-D (0.67) compared to VMAT (0.52) (P = 0.013). CK (41.86) had the largest HI compared to VMAT (26.99), HT-F (20.69), and HT-D (21.17) (P < 0.001). GI was significantly less for CK (3.96) compared to VMAT (6.76) (P = 0.001). Likewise, CK (62.4 min, 14059 MU) had the longest treatment time and MU per fraction compared to VMAT (8.5 min, 9764 MU), HT-F (13 min, 10822 MU), and HT-D (13.5 min, 11418 MU) (P < 0.001).

CONCLUSION

Both CK and HT plans achieved conformal target coverage while respecting cord tolerance. Dose heterogeneity was significantly larger in CK. VMAT required the least treatment time and MU output, but had the least steep GI, CI, and target coverage.

Для цитирования: Aljabab Saif, Vellayappan Balamurugan, Vandervoort Eric, Bahm Jamie, Zohr Robert, Sinclair John, Caudrelier Jean-Michel, Szanto Janos, Malone Shawn. Comparison of four techniques for spine stereotactic body radiotherapy: Dosimetric and efficiency analysis.. Journal of applied clinical medical physics. 2018. DOI: 10.1002/acm2.12271.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.