Prospective Clinical and Pharmacological Evaluation of the Delcath System's Second-Generation (GEN2) Hemofiltration System in Patients Undergoing Percutaneous Hepatic Perfusion with Melphalan.

de Leede Eleonora M, Burgmans Mark C, Meijer T Susanna, Martini Christian H, Tijl Fred G J, Vuyk Jaap, van Erkel Arian R, van der Velde Cornelis J H, Kapiteijn Ellen, Vahrmeijer Alexander L

Клиническое исследование II фазыСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Percutaneous hepatic perfusion (PHP) with melphalan is an effective treatment for patients with hepatic metastases, but associated with high rates of bone marrow depression. To reduce systemic toxicity, improvements have been made to the filtration system. In pre-clinical studies, the Delcath System's GEN2 filter was superior to the first-generation filters. In this clinical study, we analysed the pharmacokinetics and toxicity of PHP using the new GEN2 filter.

METHODS AND MATERIALS: Starting February 2014, two prospective phase II studies were initiated in patients with hepatic metastases from ocular melanoma or colorectal cancer. In 10 PHP procedures performed in the first 7 enrolled patients, blood samples were obtained to determine filter efficiency and systemic drug exposure. PHP was performed with melphalan 3 mg/kg with a maximum of 220 mg. Complications were assessed according to CTCAE v4.03. Response was assessed according to RECIST 1.1.

RESULTS

Pharmacokinetic analysis of blood samples showed an overall filter efficiency of 86% (range 71.1-95.5%). The mean filter efficiency decreased from 95.4% 10 min after the start of melphalan infusion to 77.5% at the end of the procedure (p = 0.051). Bone marrow depression was seen after up to 80.0% of 10 procedures, but was self-limiting and mostly asymptomatic. No hypotension-related complications or procedure-related mortality occurred.

CONCLUSION

The GEN2 filter has a higher melphalan filter efficiency compared to the first-generation filters and a more consistent performance. PHP with the GEN2 filter appears to have an acceptable safety profile, but this needs further validation in larger studies.

Для цитирования: de Leede Eleonora M, Burgmans Mark C, Meijer T Susanna, Martini Christian H, Tijl Fred G J, Vuyk Jaap, van Erkel Arian R, van der Velde Cornelis J H, Kapiteijn Ellen, Vahrmeijer Alexander L. Prospective Clinical and Pharmacological Evaluation of the Delcath System's Second-Generation (GEN2) Hemofiltration System in Patients Undergoing Percutaneous Hepatic Perfusion with Melphalan.. Cardiovascular and interventional radiology. 2017. DOI: 10.1007/s00270-017-1630-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.