Influence of simulated adverse events of the floating mass transducer in a mechanical middle ear model.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Graf Lukas, Hürzeler Nathan, Wey Lucas, Mojallal Hamidreza, Stieger Christof, Arnold Andreas
Аннотация
BACKGROUND: The Vibrant Soundbridge Implant provides multiple options for coupling the Floating Mass Transducer (FMT) to the ossicles. A parametric evaluation in clinical trials is difficult.
AIMS/OBJECTIVES: We studied the vibratory output of the FMT with three couplers-short process, former long process, and current long process-in an acoustic-mechanical middle ear model (AMEM) under simulated impaired coupling conditions.
MATERIALS AND METHODS: The AMEM, with life-sized ossicles, was tested via acoustic and electrical (FMT-) stimulation. Simulated impairments included accidental bone-contact, cable tension, and insufficient connection. Stapedial footplate vibrations were measured using Laser Doppler.
RESULTS: Acoustic stimulation generally fulfilled the ASTM standard. With normal coupling, SP- and LP-couplers performed comparable to temporal bone data. Impaired coupling led to various effects: bone-contact reduced the transmission up to 20 dB, cable tension minimally affected SP and current LP couplers but reduced the sound transmission for the former LP coupler. Connecting the SP-coupler with only two of four titanium legs caused a single frequency dip without affecting overall magnitude, whereas off-axis LP fixation reduced the output by 15-20 dB around 1000 Hz.
CONCLUSIONS AND SIGNIFICANCE: The AMEM generated reproducible, ASTM-compliant measurements. SP and current LP couplers showed resilience to potential FMT implantation impairments in most frequencies.
AIMS/OBJECTIVES: We studied the vibratory output of the FMT with three couplers-short process, former long process, and current long process-in an acoustic-mechanical middle ear model (AMEM) under simulated impaired coupling conditions.
MATERIALS AND METHODS: The AMEM, with life-sized ossicles, was tested via acoustic and electrical (FMT-) stimulation. Simulated impairments included accidental bone-contact, cable tension, and insufficient connection. Stapedial footplate vibrations were measured using Laser Doppler.
RESULTS: Acoustic stimulation generally fulfilled the ASTM standard. With normal coupling, SP- and LP-couplers performed comparable to temporal bone data. Impaired coupling led to various effects: bone-contact reduced the transmission up to 20 dB, cable tension minimally affected SP and current LP couplers but reduced the sound transmission for the former LP coupler. Connecting the SP-coupler with only two of four titanium legs caused a single frequency dip without affecting overall magnitude, whereas off-axis LP fixation reduced the output by 15-20 dB around 1000 Hz.
CONCLUSIONS AND SIGNIFICANCE: The AMEM generated reproducible, ASTM-compliant measurements. SP and current LP couplers showed resilience to potential FMT implantation impairments in most frequencies.