DOI 10.1080/00016489.2025.2451089

Медицинская библиотека · Acta oto-laryngologica · 2025

Influence of simulated adverse events of the floating mass transducer in a mechanical middle ear model.

Graf Lukas, Hürzeler Nathan, Wey Lucas, Mojallal Hamidreza, Stieger Christof, Arnold Andreas

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The Vibrant Soundbridge Implant provides multiple options for coupling the Floating Mass Transducer (FMT) to the ossicles. A parametric evaluation in clinical trials is difficult.

AIMS/

OBJECTIVES

We studied the vibratory output of the FMT with three couplers-short process, former long process, and current long process-in an acoustic-mechanical middle ear model (AMEM) under simulated impaired coupling conditions.

MATERIALS AND

METHODS

The AMEM, with life-sized ossicles, was tested via acoustic and electrical (FMT-) stimulation. Simulated impairments included accidental bone-contact, cable tension, and insufficient connection. Stapedial footplate vibrations were measured using Laser Doppler.

RESULTS

Acoustic stimulation generally fulfilled the ASTM standard. With normal coupling, SP- and LP-couplers performed comparable to temporal bone data. Impaired coupling led to various effects: bone-contact reduced the transmission up to 20 dB, cable tension minimally affected SP and current LP couplers but reduced the sound transmission for the former LP coupler. Connecting the SP-coupler with only two of four titanium legs caused a single frequency dip without affecting overall magnitude, whereas off-axis LP fixation reduced the output by 15-20 dB around 1000 Hz.

CONCLUSIONS AND SIGNIFICANCE: The AMEM generated reproducible, ASTM-compliant measurements. SP and current LP couplers showed resilience to potential FMT implantation impairments in most frequencies.

Для цитирования: Graf Lukas, Hürzeler Nathan, Wey Lucas, Mojallal Hamidreza, Stieger Christof, Arnold Andreas. Influence of simulated adverse events of the floating mass transducer in a mechanical middle ear model.. Acta oto-laryngologica. 2025. DOI: 10.1080/00016489.2025.2451089.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.