The Annals of otology, rhinology, and laryngology · 2016
Radiological and NRT-Ratio-Based Estimation of Slim Straight Cochlear Implant Electrode Positions: A Multicenter Study.
СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mittmann Philipp, Todt Ingo, Ernst Arneborg, Rademacher Grit, Mutze Sven, Göricke Sophia, Schlamann Marc, Lang Stephan, Arweiler-Harbeck Diana, Christov Florian
Аннотация
OBJECTIVES: An intraoperative neural response telemetry-ratio (NRT-ratio) was established, which can provide information about the intraoperative intracochlear electrode array position for perimodiolar electrodes.
METHODS: In a retrospective controlled study in 2 tertiary referral centers, the electrophysiological data sets of 50 patients with measured intraoperative auto-NRTs and postoperative radiological examinations were evaluated. All patients were implanted with Nucleus slim straight electrodes. The NRT-ratio was calculated by dividing the average auto-NRT data from electrodes 16 to 18 with the average from electrodes 5 to 7. Using a flat panel tomography system or a computed tomography, the position of the electrode array was certified radiological.
RESULTS: Radiologically, 2 out of 50 patients were identified with an electrode translocated from the scala tympani into the scala vestibuli. The radiologically estimated electrodes indicating a scalar change showed a regular NRT-ratio but nonspecific NRT-level changes at the localization of translocation.
METHODS: In a retrospective controlled study in 2 tertiary referral centers, the electrophysiological data sets of 50 patients with measured intraoperative auto-NRTs and postoperative radiological examinations were evaluated. All patients were implanted with Nucleus slim straight electrodes. The NRT-ratio was calculated by dividing the average auto-NRT data from electrodes 16 to 18 with the average from electrodes 5 to 7. Using a flat panel tomography system or a computed tomography, the position of the electrode array was certified radiological.
RESULTS: Radiologically, 2 out of 50 patients were identified with an electrode translocated from the scala tympani into the scala vestibuli. The radiologically estimated electrodes indicating a scalar change showed a regular NRT-ratio but nonspecific NRT-level changes at the localization of translocation.