DOI 10.1016/j.ejmp.2026.105789

Медицинская библиотека · Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026

Towards novel methods of dosimetry in pediatric and neonatal head computed tomography: Comparative study using two novel and cost effective fabricated in-house phantoms.

Sekkat H, Khallouqi A, Madkouri Y, Talbi M, Halimi A, El Rhazouani O, El Mouden O, Bannan A

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: Accurate radiation dose assessment in pediatric computed tomography (CT) is essential due to children's higher radiosensitivity and anatomical differences from adults. Standard homogeneous adult phantoms often underestimate pediatric doses, especially in head imaging. This study performs a comparative dosimetry using two novel in-house developed heterogeneous head phantoms.

METHODS

Tissue-equivalent materials (TEMs) for cranial bone, brain matter, cerebrospinal fluid (CSF), and tumor/mass were formulated using Monte Carlo (GATE) simulations, XMuDat calculations, and Hounsfield unit data from pediatric head CT scans. Organ doses were measured using optically stimulated luminescence dosimeters under clinical protocols at 100 and 120 kVp using machined inserts in the phantoms for the different TEMs.

RESULTS

Physical and radiological properties, density, electron density, effective atomic number and mass attenuation, showed close agreement between simulations and theoretical values (<9% for most tissues). CT numbers matched reference pediatric HU ranges, confirming anatomical realism. Cranial bone peripheral doses ranged from 18.48 ± 0.16 mGy (neonatal, 100 kVp) to 24.66 ± 0.04 mGy (pediatric, 120 kVp), CSF peripheral doses from 17.75 ± 0.11 mGy to 24.00 ± 0.68 mGy, brain matter from 16.09 ± 0.27 mGy to 19.95 ± 0.38 mGy. Dose increases were ∼ 18-27% between neonatal and pediatric 100 kVp protocols, and ∼ 4-8% when increasing pediatric tube potential from 100 to 120 kVp.

CONCLUSION

These phantoms provide a realistic platform for pediatric head CT dose quantification, supporting patient-size-specific optimization and establishment of local diagnostic reference levels (DRLs) based on size-specific dose estimates.

Для цитирования: Sekkat H, Khallouqi A, Madkouri Y, Talbi M, Halimi A, El Rhazouani O, El Mouden O, Bannan A. Towards novel methods of dosimetry in pediatric and neonatal head computed tomography: Comparative study using two novel and cost effective fabricated in-house phantoms.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2026. DOI: 10.1016/j.ejmp.2026.105789.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.