DOI 10.1161/circulationaha.105.597013

Медицинская библиотека · Circulation · 2006

Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations.

Sommer Torsten, Naehle Claas P, Yang Alexander, Zeijlemaker Volkert, Hackenbroch Matthias, Schmiedel Alexandra, Meyer Carsten, Strach Katharina, Skowasch Dirk, Vahlhaus Christian, Litt Harold, Schild Hans

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The purpose of the present study was to evaluate a strategy for safe performance of extrathoracic magnetic resonance imaging (MRI) in non-pacemaker-dependent patients with cardiac pacemakers.

METHODS AND

RESULTS

Inclusion criteria were presence of a cardiac pacemaker and urgent clinical need for an MRI examination. Pacemaker-dependent patients and those requiring examinations of the thoracic region were excluded. The study group consisted of 82 pacemaker patients who underwent a total of 115 MRI examinations at 1.5T. To minimize radiofrequency-related lead heating, the specific absorption rate was limited to 1.5 W/kg. All pacemakers were reprogrammed before MRI: If heart rate was <60 bpm, the asynchronous mode was programmed to avoid magnetic resonance (MR)-induced inhibition; if heart rate was >60 bpm, sense-only mode was used to avoid MR-induced competitive pacing and potential proarrhythmia. Patients were monitored with ECG and pulse oximetry. All pacemakers were interrogated immediately before and after the MRI examination and after 3 months, including measurement of pacing capture threshold (PCT) and serum troponin I levels. All MR examinations were completed safely. Inhibition of pacemaker output or induction of arrhythmias was not observed. PCT increased significantly from pre- to post-MRI (P=0.017). In 2 of 195 leads, an increase in PCT was only detected at follow-up. In 4 of 114 examinations, troponin increased from a normal baseline value to above normal after MRI, and in 1 case (troponin pre-MRI 0.02 ng/mL, post-MRI 0.16 ng/mL), this increase was associated with a significant increase in PCT.

CONCLUSIONS

Extrathoracic MRI of non-pacemaker-dependent patients can be performed with an acceptable risk-benefit ratio under controlled conditions and by taking both MR- and pacemaker-related precautions.

Для цитирования: Sommer Torsten, Naehle Claas P, Yang Alexander, Zeijlemaker Volkert, Hackenbroch Matthias, Schmiedel Alexandra, Meyer Carsten, Strach Katharina, Skowasch Dirk, Vahlhaus Christian, Litt Harold, Schild Hans. Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations.. Circulation. 2006. DOI: 10.1161/circulationaha.105.597013.