Comparison of four direct homogeneous methods for the measurement of low-density lipoprotein cholesterol.
Аннотация
BACKGROUND
The primary lipoprotein risk factor is low-density lipoprotein cholesterol (LDL-C), and medication is targeted at lowering LDL-C values. To clarify the usefulness of direct homogeneous assays for LDL-C measurement, we compared the values obtained by various reagents to those obtained by the Friedewald equation and analyzed different reactivity to IDL/VLDL and LDL.
METHODS
Serum samples were collected from 55 patients with hypercholesterolemia. The LDL-C concentrations were determined by four direct homogeneous assays using reagent A (Kyowa Medex), B (Sekisui Medical), C (Denka Seiken), and D (Sysmex), which are commercially available.
RESULTS
Significant correlation was observed in LDL-C values obtained by the homogeneous assays and the Friedewald equation. However, there were two discrepancies in reagents B and C, respectively. These assays showed 40% and 55% lower LDL-C values than those calculated by the Friedewald equation, respectively. Reactivity to the IDL fraction in reagents B and C was lower than in reagents A and D.
CONCLUSIONS
Direct homogeneous assays for LDL-C are suitable for routine laboratory examination. However, it was shown that attention should be given to the different reactivity to IDL and LDL among reagents in some clinical samples.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.