Медицинская библиотека · Clinical laboratory · 2010

Comparison of four direct homogeneous methods for the measurement of low-density lipoprotein cholesterol.

Yamada Koji, Tsuji Naoki, Fujita Tomoko, Tanaka Maki, Kuribayashi Kageaki, Kobayashi Daisuke, Watanabe Naoki

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The primary lipoprotein risk factor is low-density lipoprotein cholesterol (LDL-C), and medication is targeted at lowering LDL-C values. To clarify the usefulness of direct homogeneous assays for LDL-C measurement, we compared the values obtained by various reagents to those obtained by the Friedewald equation and analyzed different reactivity to IDL/VLDL and LDL.

METHODS

Serum samples were collected from 55 patients with hypercholesterolemia. The LDL-C concentrations were determined by four direct homogeneous assays using reagent A (Kyowa Medex), B (Sekisui Medical), C (Denka Seiken), and D (Sysmex), which are commercially available.

RESULTS

Significant correlation was observed in LDL-C values obtained by the homogeneous assays and the Friedewald equation. However, there were two discrepancies in reagents B and C, respectively. These assays showed 40% and 55% lower LDL-C values than those calculated by the Friedewald equation, respectively. Reactivity to the IDL fraction in reagents B and C was lower than in reagents A and D.

CONCLUSIONS

Direct homogeneous assays for LDL-C are suitable for routine laboratory examination. However, it was shown that attention should be given to the different reactivity to IDL and LDL among reagents in some clinical samples.

Для цитирования: Yamada Koji, Tsuji Naoki, Fujita Tomoko, Tanaka Maki, Kuribayashi Kageaki, Kobayashi Daisuke, Watanabe Naoki. Comparison of four direct homogeneous methods for the measurement of low-density lipoprotein cholesterol.. Clinical laboratory. 2010.