DOI 10.1007/s00259-020-04689-y
A kinetics-based approach to amyloid PET semi-quantification.
Аннотация
PURPOSE: To develop and validate a semi-quantification method (time-delayed ratio, TDr) applied to amyloid PET scans, based on tracer kinetics information.
METHODS
The TDr method requires two static scans per subject: one early (~ 0-10 min after the injection) and one late (typically 50-70 min or 90-100 min after the injection, depending on the tracer). High perfusion regions are delineated on the early scan and applied onto the late scan. A SUVr-like ratio is calculated between the average intensities in the high perfusion regions and the late scan hotspot. TDr was applied to a naturalistic multicenter dataset of 143 subjects acquired with [18F]florbetapir. TDr values are compared to visual evaluation, cortical-cerebellar SUVr, and to the geometrical semi-quantification method ELBA. All three methods are gauged versus the heterogeneity of the dataset.
RESULTS
TDr shows excellent agreement with respect to the binary visual assessment (AUC = 0.99) and significantly correlates with both validated semi-quantification methods, reaching a Pearson correlation coefficient of 0.86 with respect to ELBA.
CONCLUSIONS
TDr is an alternative approach to previously validated ones (SUVr and ELBA). It requires minimal image processing; it is independent on predefined regions of interest and does not require MR registration. Besides, it takes advantage on the availability of early scans which are becoming common practice while imposing a negligible added patient discomfort.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.