An evaluation of hardware and surgical complications with deep brain stimulation based on diagnosis and lead location.

Falowski Steven, Ooi Yinn Cher, Smith Adam, Verhargen Metman Leonard, Bakay Roy A E

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Deep brain stimulation is the most frequently performed neurosurgical procedure for movement disorders. This procedure is well tolerated, but not free of complications. Analysis of hardware complications based on patient diagnosis and lead location could prove valuable in recognizing potential pitfalls and patients at higher risk.

METHODS

This review analyzes the most common surgery-related complications that may occur based on diagnosis and lead location. Patients were categorized based on diagnosis - Parkinson's disease (PD), dystonia, and essential tremor (ET) - as well as by lead location - subthalamic nucleus (STN), globus pallidus interna (GPi), and ventral intermediate nucleus of the thalamus (Vim). It is a retrospective review of 326 patients undergoing 949 procedures over a 10-year period by one surgeon. Fisher's exact test and χ(2) test were employed and multivariate logistic regression analysis was performed to identify the significant variables of correlation.

RESULTS

Overall lead revision was observed at 5.7%, but was observed at 11.9% of GPi lead placements, and 10.7% of dystonia patients with only 4.6% of STN lead placements. Total extension revision was at 2.5%, but observed at 5.3% for dystonia patients and at only 1.4% for ET patients. Overall infection rate was at 1.9% with the highest rate observed in dystonia and ET patients. Postoperative complications with hardware, erosion, infection, and delayed stimulation failure were observed more often with ET and dystonia than with PD. This difference was statistically significant between dystonia and PD (p < 0.03) but not between the other disease entities (p > 0.05). On multivariate analysis, age and gender had no correlation with these complications. PD had significantly fewer complications on forward selection regression analysis (p = 0.004). Asymptomatic intracerebral hemorrhage was at 2.5% with the majority in Vim and none observed in GPi placements. There was only one symptomatic hemorrhage with a permanent deficit. Infarcts were observed at 0.8%. There were no mortalities.

CONCLUSION

This large series of patients and long-term follow-up demonstrate that risks of complications are not universal among movement disorder patients. Diagnosis and lead location are important risk stratification factors in determining complications.

Для цитирования: Falowski Steven, Ooi Yinn Cher, Smith Adam, Verhargen Metman Leonard, Bakay Roy A E. An evaluation of hardware and surgical complications with deep brain stimulation based on diagnosis and lead location.. Stereotactic and functional neurosurgery. 2012. DOI: 10.1159/000338254.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.