Accuracy of 2-hydroxyglutarate quantification by short-echo proton-MRS at 3 T: a phantom study.

Bertolino Nicola, Marchionni Chiara, Ghielmetti Francesco, Burns Brian, Finocchiaro Gaetano, Anghileri Elena, Bruzzone Maria Grazia, Minati Ludovico

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: We set out to investigate the potential confounding effect of variable concentration of N-acetyl-l-aspartate (NAA) and Glutamate (Glu) on measurement of the brain oncometabolite 2-hydroxyglutarate (2HG) using a standard MRS protocol. This issue may arise due to spectral overlap at clinical magnetic field strengths and thus complicate the usage of 2HG as a putative biomarker of gliomas bearing mutations of the isocitrate dehydrogenase (IDH) 1 and 2 genes.

METHODS

Spectra from 25 phantoms (50 mL falcon test tubes) containing a range of known concentrations of 2HG, NAA and Glu were acquired using a clinical 3 T scanner with a quadrature head coil, single-voxel point-resolved spectroscopy sequence with TE = 30 ms. Metabolite concentrations were estimated by linear combination analysis and a simulated basis set.

RESULTS

NAA and Glu concentrations can have a significant confounding effect on 2HG measurements, whereby the negative changes in concentration of these metabolites typically observed in (peri)lesional areas can lead to under-estimation of 2HG concentration with respect to spectra acquired in presence of physiological levels of NAA and Glu.

CONCLUSION

The confounding effect of NAA and Glu concentration changes needs to be considered: in patients, it may mask the presence of 2HG at low concentrations, however it is not expected to lead to false positives. 2HG data acquired using standard short echo-time MRS protocols should be considered with caution.

Для цитирования: Bertolino Nicola, Marchionni Chiara, Ghielmetti Francesco, Burns Brian, Finocchiaro Gaetano, Anghileri Elena, Bruzzone Maria Grazia, Minati Ludovico. Accuracy of 2-hydroxyglutarate quantification by short-echo proton-MRS at 3 T: a phantom study.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2014. DOI: 10.1016/j.ejmp.2014.03.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.