MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/01.ijg.0000196653.77836.af

Медицинская библиотека · Journal of glaucoma · 2006

Ab interno trabeculectomy: development of a novel device (Trabectome) and surgery for open-angle glaucoma.

Francis Brian A, See Robert F, Rao Narsing A, Minckler Don S, Baerveldt George

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To design an instrument to selectively remove trabecular meshwork and Schlemm's canal inner wall (SCIW), and demonstrate its effectiveness by histologic analysis of treated cadaveric human tissue.

METHODS

The design parameters of the instrument were the ability to permanently remove a segment of trabecular meshwork and Schlemm's canal inner wall without causing damage to surrounding tissue, and to allow use with standard anterior segment surgical techniques and equipment via an ab interno approach. Treatment was applied to 20 segments of human corneoscleral rims. The treated areas were examined using a confocal microscope and compared with matching areas in untreated controls and simulated goniotomy.

RESULTS

The resultant instrument system surgically removes the trabecular meshwork and Schlemm's canal inner wall from an anterior chamber approach. It consists of a disposable surgical handpiece with irrigation, aspiration, and electrocautery to focally ablate the target tissues. The attached console includes a high-frequency (550 KHz) electrosurgical generator and irrigation/aspiration controlled by a foot pedal. Histologic examination of specimens treated with the Trabectome displayed disruption of the trabecular meshwork and Schlemm's canal inner wall without damage to surrounding structures. The specimens treated by simulated goniotomy displayed significant damage to the outer wall of Schlemm's canal and the surrounding sclera. The controls showed no disruption or damage to any tissues.

CONCLUSIONS

The Trabectome system is designed for performing trabeculectomy via an ab interno approach. It successfully removed sections of trabecular meshwork and Schlemm's canal inner wall with less injury to the adjacent tissue compared with goniotomy knife in vitro. Theoretically, this procedure should provide direct access of aqueous humor to Schlemm's canal.

Для цитирования: Francis Brian A, See Robert F, Rao Narsing A, Minckler Don S, Baerveldt George. Ab interno trabeculectomy: development of a novel device (Trabectome) and surgery for open-angle glaucoma.. Journal of glaucoma. 2006. DOI: 10.1097/01.ijg.0000196653.77836.af.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.