In vivo dosimetry of thyroid doses from different irradiated sites in children and adolescents: a cross-sectional study.
Аннотация
BACKGROUND
Scattered radiation can be assessed by in vivo dosimetry. Thyroid tissue is sensitive to radiation, even at doses <10 cGy. This study compared the scattered dose to the thyroid measured by thermoluminescent dosimeters (TLDs) and the estimated one by treatment planning system (TPS).
METHODS
During radiotherapy to sites other than the thyroid of 16 children and adolescents, seventy-two TLD measurements at the thyroid were compared with TPS estimation.
RESULTS
The overall TPS/TLD bias was 1.02 (95% LA 0.05 to 21.09). When bias was stratified by treatment field, the TPS overestimated TLD values at doses <1 cGy and underestimated them at doses >10 cGy. The greatest bias was found in pelvis and abdomen: 15.01 (95% LA 9.16 to 24.61) and 5.12 (95% LA 3.04 to 8.63) respectively. There was good agreement in orbit, head, and spine: bias 1.52 (95% LA 0.48 to 4.79), 0.44 (95% LA 0.11 to 1.82) and 0.83 (0.39 to 1.76) respectively. There was small agreement with broad limits for lung and mediastinum: 1.13 (95% LA 0.03 to 40.90) and 0.39 (95% LA 0.02 to 7.14) respectively.
CONCLUSIONS
The scattered dose can be measured with TLDs, and TPS algorithms for outside structures should be improved.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.