Frameless robotically targeted stereotactic brain biopsy: feasibility, diagnostic yield, and safety.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Bekelis Kimon, Radwan Tarek A, Desai Atman, Roberts David W

Аннотация

OBJECT: Frameless stereotactic brain biopsy has become an established procedure in many neurosurgical centers worldwide. Robotic modifications of image-guided frameless stereotaxy hold promise for making these procedures safer, more effective, and more efficient. The authors hypothesized that robotic brain biopsy is a safe, accurate procedure, with a high diagnostic yield and a safety profile comparable to other stereotactic biopsy methods.

METHODS: This retrospective study included 41 patients undergoing frameless stereotactic brain biopsy of lesions (mean size 2.9 cm) for diagnostic purposes. All patients underwent image-guided, robotic biopsy in which the SurgiScope system was used in conjunction with scalp fiducial markers and a preoperatively selected target and trajectory. Forty-five procedures, with 50 supratentorial targets selected, were performed.

RESULTS: The mean operative time was 44.6 minutes for the robotic biopsy procedures. This decreased over the second half of the study by 37%, from 54.7 to 34.5 minutes (p < 0.025). The diagnostic yield was 97.8% per procedure, with a second procedure being diagnostic in the single nondiagnostic case. Complications included one transient worsening of a preexisting deficit (2%) and another deficit that was permanent (2%). There were no infections.

CONCLUSIONS: Robotic biopsy involving a preselected target and trajectory is safe, accurate, efficient, and comparable to other procedures employing either frame-based stereotaxy or frameless, nonrobotic stereotaxy. It permits biopsy in all patients, including those with small target lesions. Robotic biopsy planning facilitates careful preoperative study and optimization of needle trajectory to avoid sulcal vessels, bridging veins, and ventricular penetration.