Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine · 2005
Incidence of cardiac arrhythmias with therapeutic versus diagnostic ultrasound and intravenous microbubbles.
СтатьяПоддержка NIH: внешняя программаПоддержка службы здравоохранения СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Chapman Scott, Windle John, Xie Feng, McGrain Anna, Porter Thomas R
Аннотация
OBJECTIVE: The purpose of this study was to determine the type of arrhythmias induced with therapeutic versus diagnostic transthoracic low-frequency ultrasound (TLFUS) transducers in the presence of intravenous microbubbles.
METHODS: Intravenous perfluorocarbon-exposed sonicated dextrose albumin (PESDA) microbubbles were infused or given as a bolus injection while TLFUS was applied in the standard parasternal and apical views with either a 1-MHz therapeutic ultrasound transducer or high-mechanical-index diagnostic ultrasound (1.7 MHz).
RESULTS: Significantly more ectopy was produced by the therapeutic transducer, especially at higher-intensity settings in the continuous wave mode after bolus injections of PESDA (P < .001 compared with lower intensities and lower continuous infusion rates). Six patients (15%) had either clinical supraventricular tachycardia or nonsustained ventricular tachycardia after intravenous PESDA with therapeutic TLFUS. In comparison, diagnostic high-mechanical-index ultrasound produced only isolated ventricular ectopy and no sustained ventricular arrhythmias.
CONCLUSIONS: Intravenously injected microbubbles and low-frequency therapeutic transducers operating at longer duty cycles and wide beam widths have the capability of eliciting clinically important arrhythmias in patients at high risk for such events.
METHODS: Intravenous perfluorocarbon-exposed sonicated dextrose albumin (PESDA) microbubbles were infused or given as a bolus injection while TLFUS was applied in the standard parasternal and apical views with either a 1-MHz therapeutic ultrasound transducer or high-mechanical-index diagnostic ultrasound (1.7 MHz).
RESULTS: Significantly more ectopy was produced by the therapeutic transducer, especially at higher-intensity settings in the continuous wave mode after bolus injections of PESDA (P < .001 compared with lower intensities and lower continuous infusion rates). Six patients (15%) had either clinical supraventricular tachycardia or nonsustained ventricular tachycardia after intravenous PESDA with therapeutic TLFUS. In comparison, diagnostic high-mechanical-index ultrasound produced only isolated ventricular ectopy and no sustained ventricular arrhythmias.
CONCLUSIONS: Intravenously injected microbubbles and low-frequency therapeutic transducers operating at longer duty cycles and wide beam widths have the capability of eliciting clinically important arrhythmias in patients at high risk for such events.