DOI 10.1371/journal.pone.0243737

Медицинская библиотека · PloS one · 2020

Assessment of the sublingual microcirculation with the GlycoCheck system: Reproducibility and examination conditions.

Eickhoff Mie Klessen, Winther Signe Abitz, Hansen Tine Willum, Diaz Lars Jorge, Persson Frederik, Rossing Peter, Frimodt-Møller Marie

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The glycocalyx is an extracellular layer lining the lumen of the vascular endothelium, protecting the endothelium from shear stress and atherosclerosis and contributes to coagulation, immune response and microvascular perfusion. The GlycoCheck system estimates glycocalyx' thickness in vessels under the tongue from perfused boundary region (PBR) and microvascular perfusion (red blood cell (RBC) filling) via a camera and dedicated software.

OBJECTIVES

Evaluating reproducibility and influence of examination conditions on measurements with the GlycoCheck system.

METHODS

Open, randomised, controlled study including 42 healthy smokers investigating day-to-day, side-of-tongue, inter-investigator variance, intraclass-correlation (ICC) and influence of examination conditions at intervals from 0-180 minutes on PBR and RBC filling.

RESULTS

Mean (SD) age was 24.9 (6.1) years, 52% were male. There was no significant intra- or inter-investigator variation for PBR or RBC filling nor for PBR for side-of-tongue. A small day-to-day variance was found for PBR (0.012μm, p = 0.007) and RBC filling (0.003%, p = 0.005) and side-of-tongue, RBC filling (0.025%, p = 0.009). ICC was modest but highly improved by increasing measurements. Small significant influence of cigarette smoking (from 40-180 minutes), high calorie meal intake and coffee consumption was found. The latter two peaking immediately and tapering off but remained significant up to 180 minutes, highest PBR changes for the three being 0.042μm (p<0.05), 0.183μm (p<0.001) and 0.160μm (p<0.05) respectively.

CONCLUSIONS

Measurements with the GlycoCheck system have a moderate reproducibility, but highly increases with multiple measurements and a small day-to-day variability. Smoking, meal and coffee intake had effects up to 180 minutes, abstinence is recommended at least 180 minutes before GlycoCheck measurements. Future studies should standardise conditions during measurements.

Для цитирования: Eickhoff Mie Klessen, Winther Signe Abitz, Hansen Tine Willum, Diaz Lars Jorge, Persson Frederik, Rossing Peter, Frimodt-Møller Marie. Assessment of the sublingual microcirculation with the GlycoCheck system: Reproducibility and examination conditions.. PloS one. 2020. DOI: 10.1371/journal.pone.0243737.

Предметные рубрики MeSH