MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ejmp.2025.104921

Медицинская библиотека · Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2025

Determining the optimal surface tracking area on the stereotactic mask in brain stereotactic radiosurgery using thermo-optical surface-guided radiotherapy.

Katayama Hiroki, Takahashi Yosuke, Kitaoka Motonori, Kawasaki Hiroki, Tanii Takashi, Taniguchi Yayoi, Tsuzuki Masato

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Surface-guided radiotherapy (SGRT) using ExacTrac Dynamic (EXTD) combines the optical surface with "thermal" imaging. We investigated the relationship between surface temperature changes and position errors detected by EXTD and identified the optimal surface tracking area (STA) on a stereotactic mask for brain stereotactic radiosurgery.

METHODS

A phantom with a heat pad and a stereotactic mask over it was set up on the linac, and its surface was assigned to the STA. After powering on the heat pad, we investigated the change in the mask's surface temperature for a detected 1.0-mm position error. Subsequently, stereotactic masks were created for six volunteers. The temperatures of the forehead, nose, mouth, and cheek areas on the mask were measured using a thermography camera. We identified the temperature stabilization areas on the mask during the treatment.

RESULTS

Position errors were detected to be 1.0 mm when the surface temperature increased by approximately 0.5 °C in the phantom study. Next, the average temperature stabilization time on the mask was 4.2, 3.7, 2.3 and 1.8 min in the forehead, nose, mouth, and cheeks, respectively. The surface temperature of the mask stabilized after 4 min in the nose, mouth, and cheeks, except for the forehead area. However, the mask temperature of the nose and mouth areas decreased with the breathing (>1.0 °C) in five of the six volunteers.

CONCLUSION

Assigning the STA to areas exhibiting temperature stability (<0.5 °C) is crucial, and we recommend assigning the STA to the cheek area after fitting the stereotactic mask for 4 min.

Для цитирования: Katayama Hiroki, Takahashi Yosuke, Kitaoka Motonori, Kawasaki Hiroki, Tanii Takashi, Taniguchi Yayoi, Tsuzuki Masato. Determining the optimal surface tracking area on the stereotactic mask in brain stereotactic radiosurgery using thermo-optical surface-guided radiotherapy.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2025. DOI: 10.1016/j.ejmp.2025.104921.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.