Preliminary dosimetric comparison between fixed and rotating source stereotactic radiosurgery systems.

Draeger Emily, Chen Zhe Jay, Hansen James E, Chiang Veronica, Tien Christopher J

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The Akesis Galaxy RTi (AK) is a novel rotational 60 Co-based cranial stereotactic radiosurgery (SRS) system. While similar systems have been compared against the fixed-source Leksell Gamma Knife (GK) system using stylized phantoms, dosimetric plan quality with realistic anatomy has yet to be characterized for this or any other rotating system versus GK. This study aims to benchmark AK dosimetric performance against GK by retrospectively replanning previously-treated GK patients at our institution.

METHODS

Thirteen patients, previously treated on a GK Icon, were re-planned on the AK treatment planning system using the same prescription doses and isodoses as the original GK plans. The cohort includes patients treated for brain metastases, schwannomas, pituitary adenomas, trigeminal neuralgias, and arteriovenous malformations. Plans are evaluated with target coverage metrics (Dmin , Dmean , D95% , V150% ) and dose conformality indices: Radiation Therapy Oncology Group conformity index (CI), selectivity, Paddick CI (PCI), gradient index (GI).

RESULTS

AK plans use fewer shots and larger collimation compared to GK plans, resulting in statistically significant reductions in treatment time (p = 0.047) by as much as 88.4 minutes while maintaining comparable target V100% . For most metastatic cases, GK produces higher Dmin (16.0-25.9 vs. 12.5-24.3 Gy, p = 0.008) while AK produces higher V150% (0.03-14.92 vs. 0.02-11.59 cc, p = 0.028). For non-metastatic cases, GK provides superior CI (p = 0.025) and GI (p = 0.044). No statistically significant differences were found in the remaining metrics.

CONCLUSION

This cohort demonstrates that the AK system is able to achieve largely comparable dosimetric results to GK, typically with shorter treatment times. Further investigation with a larger cohort is underway.

Для цитирования: Draeger Emily, Chen Zhe Jay, Hansen James E, Chiang Veronica, Tien Christopher J. Preliminary dosimetric comparison between fixed and rotating source stereotactic radiosurgery systems.. Journal of applied clinical medical physics. 2023. DOI: 10.1002/acm2.13907.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.