Dosimetric evaluation in Helical TomoTherapy for lung SBRT using Monte Carlo-based independent dose verification software.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kosaka Takahiro, Takatsu Jun, Inoue Tatsuya, Iijima Kotaro, Suzuki Michimasa, Murakami Naoya, Shikama Naoto

Аннотация

PURPOSE: To elucidate the dosimetric errors caused by a model-based algorithm in lung stereotactic body radiation therapy (SBRT) with Helical TomoTherapy (HT) using Monte Carlo (MC)-based dose verification software.

METHODS: For 38 plans of lung SBRT, the dose calculation accuracy of a treatment planning system (TPS) of HT was compared with the results of DoseCHECK, the commercial MC-based independent verification software. The following indices were extracted to evaluate the correlation of dosimetric errors: (1) target volume, (2) average computed tomography (CT) value of the planning target volume (PTV) margin, and (3) average CT value of surrounding 2-mm area of the PTV (PTV ring). Receiver operating characteristic (ROC) analyses determined the threshold for 5% of differences in PTV D95%. Then, the 38 plans were classified into two groups using the cutoff values of ROC analysis for these three indices. Dosimetric differences between groups were statistically compared using the Mann-Whitney U test.

RESULTS: TPS of HT overestimated by more than 5% in the PTV D95% in 16 of 38 plans. The PTV ring showed the strongest correlation with dosimetric differences. The cutoff value for the target volume, the PTV margin, and the PTV ring was 14.7 cc, -754 HU, and -708 HU, respectively. The area under the curve (AUC) for the target volume, the PTV margin, and the PTV ring were 0.835, 0.878, and 0.932, respectively. Dosimetric errors more than 5% were observed when the PTV volume was less than 15 cc or when the CT value around the target was less than -700 HU.

CONCLUSION: The TPS of HT might overestimate the PTV dose by more than 5% if any the three indices in this study were below threshold. Therefore, independent verification with an MC-based algorithm should be strongly recommended for lung SBRT in HT.