MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1167/iovs.11-8137

Медицинская библиотека · Investigative ophthalmology & visual science · 2011

Miniaturized high-intensity focused ultrasound device in patients with glaucoma: a clinical pilot study.

Aptel Florent, Charrel Thomas, Lafon Cyril, Romano Fabrice, Chapelon Jean-Yves, Blumen-Ohana Esther, Nordmann Jean-Philippe, Denis Philippe

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the relative safety and potential efficacy of high-intensity focused ultrasound cyclocoagulation by a miniaturized annular device containing six piezoceramic transducers in patients with refractory glaucoma.

METHODS

This was a three-center prospective interventional pilot study. Twelve eyes of 12 patients with refractory glaucoma were insonified using a ring-shaped probe containing six miniaturized high-frequency transducers operating at 21 MHz. Ultrasound biomicroscopy (UBM) and a complete ophthalmic examination were performed before the procedure and at 1 day, 1 week, 1 month, and 3 months after the procedure. Additional visits were performed 6 and 12 months after the procedure.

RESULTS

Intraocular pressure was significantly reduced (P < 0.01) from a mean preoperative value of 37.9 ± 10.7 mm Hg to a mean postoperative value of 27.3 ± 12.4, 25.2 ± 11.3, 25.2 ± 7.7, 24.8 ± 9.8, and 26.3 ± 5.1 mm Hg at 1 day, 1 week, 1 month, 3 months, and 6 months, respectively, and to a mean value of 24.7 ± 8.5 at the last follow-up visit. No major intraoperative or postoperative complications occurred. Minor postoperative corneal complications developed in four patients with previous corneal abnormalities: superficial punctate keratitis (n = 3) and central superficial corneal ulceration (n = 1). UBM showed cystic involution of the ciliary body in 9 of the 12 eyes and a suprachoroidal fluid space in 8 of the 12 eyes.

CONCLUSIONS

Ultrasonic circular cyclocoagulation using high-intensity focused ultrasound delivered by a circular miniaturized device containing six piezoceramic transducers seems to be an effective and well-tolerated method to reduce intraocular pressure in patients with refractory glaucoma.

Для цитирования: Aptel Florent, Charrel Thomas, Lafon Cyril, Romano Fabrice, Chapelon Jean-Yves, Blumen-Ohana Esther, Nordmann Jean-Philippe, Denis Philippe. Miniaturized high-intensity focused ultrasound device in patients with glaucoma: a clinical pilot study.. Investigative ophthalmology & visual science. 2011. DOI: 10.1167/iovs.11-8137.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.