Guigou Caroline, Hussain Raabid, Lalande Alain, Bozorg Grayeli Alexis
Аннотация
HYPOTHESIS: Transmodiolar auditory implantation via the middle ear cavity could be possible using an augmented reality system (ARS).
BACKGROUND: There is no clear landmark to indicate the cochlear apex or the modiolar axis. The ARS seems to be a promising tool for transmodiolar implantation by combining information from the preprocedure computed tomography scan (CT-scan) images to the real-time video of the surgical field.
METHODS: Eight human temporal bone resin models were included (five adults and three children). The procedure started by the identification of the modiolar axis on the preprocedure CT-scan followed by a 3D reconstruction of the images. Information on modiolar location and navigational guidance was supplemented to the reconstructed model, which was then registered with the surgical video using a point-based approach. Relative movements between the phantom and the microscope were tracked using image feature-based motion tracking. Based on the information provided via the ARS, the surgeon implanted the electrode-array inside the modiolus after drilling the helicothrema. Postprocedure CT-scan images were acquired to evaluate the registration error and the implantation accuracy.
RESULTS: The implantation could be conducted in all cases with a 2D registration error of 0.4 ± 0.24 mm. The mean entry point error was 0.6 ± 1.00 mm and the implant angular error 13.5 ± 8.93 degrees (n = 8), compatible with the procedure requirements.
CONCLUSION: We developed an image-based ARS to identify the extremities and the axis of the cochlear modiolus on intraprocedure videos. The system yielded submillimetric accuracy for implantation and remained stable throughout the experimental study.
BACKGROUND: There is no clear landmark to indicate the cochlear apex or the modiolar axis. The ARS seems to be a promising tool for transmodiolar implantation by combining information from the preprocedure computed tomography scan (CT-scan) images to the real-time video of the surgical field.
METHODS: Eight human temporal bone resin models were included (five adults and three children). The procedure started by the identification of the modiolar axis on the preprocedure CT-scan followed by a 3D reconstruction of the images. Information on modiolar location and navigational guidance was supplemented to the reconstructed model, which was then registered with the surgical video using a point-based approach. Relative movements between the phantom and the microscope were tracked using image feature-based motion tracking. Based on the information provided via the ARS, the surgeon implanted the electrode-array inside the modiolus after drilling the helicothrema. Postprocedure CT-scan images were acquired to evaluate the registration error and the implantation accuracy.
RESULTS: The implantation could be conducted in all cases with a 2D registration error of 0.4 ± 0.24 mm. The mean entry point error was 0.6 ± 1.00 mm and the implant angular error 13.5 ± 8.93 degrees (n = 8), compatible with the procedure requirements.
CONCLUSION: We developed an image-based ARS to identify the extremities and the axis of the cochlear modiolus on intraprocedure videos. The system yielded submillimetric accuracy for implantation and remained stable throughout the experimental study.